Title: A multileaf collimator phantom for the quality assurance of radiation therapy planning systems and CT simulators

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [2];  [3]
  1. Department of Medical Biophysics, University of Western Ontario, London Regional Cancer Centre, London, Ontario (Canada)
  2. Department of Medical Biophysics, University of Western Ontario, London Regional Cancer Centre, London, Ontario (Canada) and Department of Oncology, University of Western Ontario, London Regional Cancer Centre, London, Ontario (Canada)
  3. Deparment of Medical Biophysics, University of Western Ontario, London Regional Cancer Centre, London, Ontario (Canada) and Department of Oncology, University of Western Ontario, London Regional Cancer Centre, London, Ontario (Canada)

Purpose: The evolution of three-dimensional conformal radiation treatment has led to the use of multileaf collimators (MLCs) in intensity-modulated radiation therapy (IMRT) and other treatment techniques to increase the conformity of the dose distribution. A new quality assurance (QA) phantom has been designed to check the handling of MLC settings in treatment planning and delivery. Methods and materials: The phantom consists of a Perspex block with stepped edges that can be rotated in all planes. The design allows for the assessment of several MLC and micro-MLC types from various manufacturers, and is therefore applicable to most radiation therapy institutions employing MLCs. The phantom is computed tomography (CT) scanned as is a patient, and QA assessments can be made of field edge display for a variety of shapes and orientations on both radiation treatment planning systems (RTPS) and computed tomography simulators. Results: The dimensions of the phantom were verified to be physically correct within an uncertainty range of 0-0.7 mm. Errors in leaf position larger than 1 mm were easily identified by multiple observers. Conclusions: The MLC geometry phantom is a useful tool in the QA of radiation therapy with application to RTPS, CT simulators, and virtual simulation packages with MLC display capabilities.

OSTI ID:
20630953
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 60, Issue 3; Other Information: DOI: 10.1016/j.ijrobp.2004.06.013; PII: S0360-3016(04)01037-5; Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016