Title: A source of PCB contamination in modified high-volume air samplers

Journal Article · · Environmental Science and Technology
DOI:https://doi.org/10.1021/es990841+· OSTI ID:20026695

Modified Anderson High Volume (Hi-Vol) air samplers are widely used for the collection of semi-volatile organic compounds (such as PCBs) from air. The foam gasket near the main air flow path in these samplers can become contaminated with PCBs if the sampler or the gasket is stored at a location with high indoor air PCB levels. Once the gasket is contaminated, it releases PCBs back into the air stream during sampling, and as a result, incorrectly high air PCB concentrations are measured. This paper presents data demonstrating this contamination problem using measurements from two Integrated Atmospheric Deposition Network sites: one at Sleeping Bear Dunes on Lake Michigan and the other at Point Petre on Lake Ontario. The authors recommend that these gaskets be replaced by Teflon tape and that the storage history of each sampler be carefully tracked.

Research Organization:
Indiana Univ., Bloomington, IN (US)
OSTI ID:
20026695
Journal Information:
Environmental Science and Technology, Vol. 34, Issue 3; Other Information: PBD: 1 Feb 2000; ISSN 0013-936X
Country of Publication:
United States
Language:
English