Performance and electret charge of N95 respirators after decontamination [Slides]
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
- Univ. of New Mexico, Albuquerque, NM (United States)
Filtration, pressure drop and quantitative fit of N95 respirators were robust to several decontamination methods including vaporous hydrogen peroxide, wet heat, bleach, and ultraviolet light. Bleach may not have penetrated the hydrophobic outer layers of the N95 respirator. Isopropyl alcohol and detergent both severely degraded the electrostatic charge of the electret filtration layer. First data in N95 respirators that the loss of filtration efficiency was directly correlated with loss of surface potential on the filtration layer. The pressure drop was unchanged, so loss of filtration efficacy would not be apparent during a user seal check. Mechanical straps degrade with repeated mechanical cycling during extended use. Decontamination did not appear to degrade the elastic straps. Significant loss of strap elasticity would be apparent during a user negative pressure seal check.
- Research Organization:
- Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA); USDOE Laboratory Directed Research and Development (LDRD) Program
- DOE Contract Number:
- NA0003525
- OSTI ID:
- 2001725
- Report Number(s):
- SAND2022--1497C; 703360
- Country of Publication:
- United States
- Language:
- English
Similar Records
COVID-19 global pandemic planning: Performance and electret charge of N95 respirators after recommended decontamination methods
Can N95 Respirators Be Reused after Disinfection? How Many Times?
Journal Article
·
Tue Dec 15 19:00:00 EST 2020
· Experimental Biology and Medicine
·
OSTI ID:1815497
Can N95 Respirators Be Reused after Disinfection? How Many Times?
Journal Article
·
Mon May 04 20:00:00 EDT 2020
· ACS Nano
·
OSTI ID:1638355