Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Compensation for non-uniform attenuation in SPECT brain imaging

Journal Article · · Journal of Nuclear Medicine
OSTI ID:198038
; ;  [1]
  1. Univ. of Massachusetts Medical Center, Worcester, MA (United States); and others

Photon attenuation is a major limitation in performing quantitative SPECT brain imaging. A number of methods have been proposed for compensation of attenuation in regions of the body that can be modelled as a uniform attenuator. The magnitude of the errors introduced into reconstructed brain images by assuming the head to be a uniform attenuator are uncertain (the skull, sinus cavities and head holder all have different attenuation properties than brain tissue). Brain imaging is unique in that the radioisotope, for the most part, is taken up within a uniform attenuation medium (i.e., brain tissue) which is surrounded by bone (i.e., the skull) of a different density. Using this observation, Bellini`s method for attenuation compensation (which is an exact solution to the exponential Radon transform) has been modified to account for the different attenuation properties of the skull. To test this modified Bellini method, a simple mathematical phantom was designed to model the brain and a skull of varying thickness less than 7.5 mm. To model brain imaging with Tc-99m HMPAO, the attenuation coefficient of the brain tissue and skull were set to 0.15 cm{sup -1} and 0.22 cm{sup -1} respectively. A ray-driven projector which accounted for non-uniform attenuation was used to simulate projection data from 128 views. The detector response and scatter were not simulated. It was observed that reconstructions processed with uniform attenuation compensation (i.e., where it was assumed that the brain tissue and the skull had the same attenuation coefficient) provided errors of 6-20%, whereas those processed with the non-uniform Bellini algorithm were biased by only 0-5%.

OSTI ID:
198038
Report Number(s):
CONF-940605--
Journal Information:
Journal of Nuclear Medicine, Journal Name: Journal of Nuclear Medicine Journal Issue: Suppl.5 Vol. 35; ISSN JNMEAQ; ISSN 0161-5505
Country of Publication:
United States
Language:
English

Similar Records

Compensation for nonuniform attenuation in SPECT brain imaging
Journal Article · Sun Mar 31 23:00:00 EST 1996 · IEEE Transactions on Nuclear Science · OSTI ID:248095

Attenuation compensation in TC-99M SPECT brain imaging: Use of attenuation maps derived from tranmission versus emission data
Journal Article · Sun May 01 00:00:00 EDT 1994 · Journal of Nuclear Medicine · OSTI ID:198042

A direct measurement of skull attenuation for quantitative SPECT
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5895346