Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Characterization of Non-uniformities in Superconducting Tunnel Junction Radiation Detectors

Journal Article · · Journal of Low Temperature Physics
 [1];  [2];  [3];  [4];  [1];  [1];  [1];  [2];  [2];  [2];  [5];  [5];  [6];  [6]
  1. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
  2. Colorado School of Mines, Golden, CO (United States)
  3. Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
  4. Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); STAR Cryoelectronics, Santa Fe, NM (United States)
  5. XIA, LLC, Oakland, CA (United States)
  6. STAR Cryoelectronics, Santa Fe, NM (United States)
The small dimensions of superconducting tunnel junction (STJ) radiation detectors have motivated the development of STJ arrays to increase detection efficiency and statistical accuracy. We have used a pulsed UV laser and X-rays to characterize and understand non-uniformities in the response of different pixels in Ta-based STJ arrays. To separate the contributions of the detector and the readout electronics, we have also injected the pulses from an arbitrary waveform generator into the array’s amplifier chain. We discuss the magnitude and the sources of the non-uniformity and approaches to reduce it.
Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA (United States); Pacific Northwest National Laboratory (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE Laboratory Directed Research and Development (LDRD) Program; USDOE National Nuclear Security Administration (NNSA); USDOE Office of Science (SC)
Grant/Contract Number:
AC05-76RL01830; AC52-07NA27344; SC0002256; SC0006214; SC0021245
OSTI ID:
1901868
Report Number(s):
PNNL-SA-179572
Journal Information:
Journal of Low Temperature Physics, Journal Name: Journal of Low Temperature Physics Journal Issue: 5-6 Vol. 209; ISSN 0022-2291
Publisher:
SpringerCopyright Statement
Country of Publication:
United States
Language:
English

References (10)

Development of Ta-based STJ X-ray Detector Arrays for Synchrotron Science journal May 2014
Non-linearities in Superconducting Tunnel Junction Radiation Detectors and Their MCA Readout journal February 2020
High density processing electronics for superconducting tunnel junction x-ray detector arrays
  • Warburton, W. K.; Harris, J. T.; Friedrich, S.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 784 https://doi.org/10.1016/j.nima.2015.02.004
journal June 2015
Energy‐resolving superconducting x‐ray detectors with charge amplification due to multiple quasiparticle tunneling journal November 1993
Multiple-tunneling noise in superconducting tunnel junctions from partial current integration journal December 2001
Vortex trapping in superconducting tunnel junction detectors conference January 2002
Accurate measurement of the first excited nuclear state in U 235 journal May 2018
Limits on the Existence of sub-MeV Sterile Neutrinos from the Decay of Be 7 in Superconducting Quantum Sensors journal January 2021
Development of Ta-Based Superconducting Tunnel Junction X-Ray Detector Arrays journal June 2013
Evaluation of Multi-Channel ADCs for Gamma-Ray Spectroscopy journal April 2013