Application of radiochromic gel dosimetry to commissioning of a megavoltage research linear accelerator for small–field animal irradiation studies

Journal Article · · Medical Physics
DOI:https://doi.org/10.1002/mp.14685· OSTI ID:1783602
 [1];  [2];  [3];  [3];  [3];  [3];  [3];  [3];  [3];  [2]
  1. Univ. of Michigan, Ann Arbor, MI (United States); University of Michigan
  2. Univ. of Michigan, Ann Arbor, MI (United States); H. Lee Moffitt Cancer Center and Research Inst., Tampa, FL (United States)
  3. Univ. of Michigan, Ann Arbor, MI (United States)
To develop and implement an efficient and accurate commissioning procedure for small-field static beam animal irradiation studies on an MV research linear accelerator (Linatron-M9) using radiochromic gel dosimetry. The research linear accelerator (Linatron-M9) is a 9 MV linac with a static fixed collimator opening of 5.08 cm diameter. Lead collimators were manually placed to create smaller fields of 2 × 2 cm2, 1 × 1 cm2, and 0.5 × 0.5 cm2. Relative dosimetry measurements were performed, including profiles, percent depth dose (PDD) curves, beam divergence, and relative output factors using various dosimetry tools, including a small volume ionization chamber (A14), GAFCHROMIC™ EBT3 film, and Clearview gel dosimeters. The gel dosimeter was used to provide a 3D volumetric reference of the irradiated fields. The Linatron profiles and relative output factors were extracted at a reference depth of 2 cm with the output factor measured relative to the 2 × 2 cm2 reference field. Absolute dosimetry was performed using A14 ionization chamber measurements, which were verified using a national standards laboratory remote dosimetry service. Absolute dosimetry measurements were confirmed within 1.4% (k = 2, 95% confidence = 5%). The relative output factor of the small fields measured with films and gels agreed with a maximum relative percent error difference between the two methods of 1.1 % for the 1 × 1 cm2 field and 4.3 % for the 0.5 × 0.5 cm2 field. These relative errors were primarily due to the variability in the collimator positioning. The measured beam profiles demonstrated excellent agreement for beam size (measured as FWHM), within approximately 0.8 mm (or less). Film measurements were more accurate in the penumbra region due to the film’s finer resolution compared with the gel dosimeter. Following the van Dyk criteria, the PDD values of the film and gel measurements agree within 11% in the buildup region starting from 0.5 cm depth and within 2.6 % beyond maximum dose and into the fall-off region for depths up to 5 cm. The 2D beam profile isodose lines agree within 0.5 mm in all regions for the 0.5 × 0.5 cm2 and the 1 × 1 cm2 fields and within 1 mm for the larger field of 2 × 2 cm2. The 2D PDD curves agree within approximately 2% of the maximum in the typical therapy region (1–4 cm) for the 1 × 1 cm2 and 2 × 2 cm2 and within 5% for the 0.5 × 0.5 cm2 field. This work provides a commissioning process to measure the beam characteristics of a fixed beam MV accelerator with detailed dosimetric evaluation for its implementation in megavoltage small animal irradiation studies. Radiochromic gel dosimeters are efficient small-field relative dosimetry tools providing 3D dose measurements allowing for full representation of dose, dosimeter misalignment corrections and high reproducibility with low inter-dosimeter variability. Altogether, radiochromic gels are valuable for fast, full relative dosimetry commissioning in comparison to films for application in high-energy small-field animal irradiation studies.
Research Organization:
Univ. of Michigan, Ann Arbor, MI (United States)
Sponsoring Organization:
National Institutes of Health; US Department of Homeland Security, Countering Weapons of Mass Destruction Office, Academic Research Initiative; USDOE National Nuclear Security Administration (NNSA), Office of Nonproliferation and Verification Research and Development
Grant/Contract Number:
NA0003920
OSTI ID:
1783602
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 3 Vol. 48; ISSN 0094-2405
Publisher:
American Association of Physicists in MedicineCopyright Statement
Country of Publication:
United States
Language:
English

References (45)

Validation of GPU-accelerated superposition-convolution dose computations for the Small Animal Radiation Research Platform journal April 2018
Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS‐483, the IAEA–AAPM international Code of Practice for reference and relative dose determination journal October 2018
Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system journal April 2019
Characterisation of two new radiochromic gel dosimeters TruView™ and ClearView™ in combination with the vista™ optical CT scanner: A feasibility study journal August 2018
High-Resolution, Small Animal Radiation Research Platform With X-Ray Tomographic Guidance Capabilities journal August 2008
Tumor Cell Response to Synchrotron Microbeam Radiation Therapy Differs Markedly From Cells in Normal Tissues journal July 2010
Development and Implementation of an End-To-End Test for Absolute Dose Verification of Small Animal Preclinical Irradiation Research Platforms journal July 2020
ESTRO ACROP: Technology for precision small animal radiotherapy research: Optimal use and challenges journal March 2018
Gel Dosimeter for Depth‐Dose Measurements journal May 1957
Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging journal October 1984
Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gel journal December 1996
Absorption spectra variations of EBT radiochromic film from radiation exposure journal June 2005
The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels journal January 2006
Dose integration characteristics in normoxic polymer gel dosimetry investigated using sequential beam irradiation journal July 2007
A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform journal August 2009
The evaluation of 6 and 18 MeV electron beams for small animal irradiation journal September 2009
Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/ optical-CT 3D dosimetry system journal February 2010
Polymer gel dosimetry journal February 2010
Small animal radiotherapy research platforms journal May 2011
Cone-beam optical computed tomography for gel dosimetry II: imaging protocols journal February 2011
Dosimetric characterization of an image-guided stereotactic small animal irradiator journal March 2011
Evaluation of radiochromic gel dosimetry and polymer gel dosimetry in a clinical dose verification journal August 2013
Considerations for x-ray CT polymer gel dosimetry journal June 2013
Clinical Applications of Gel Dosimeters journal December 2006
True 3D chemical dosimetry (gels, plastics): Development and clinical role journal January 2015
Radiochromic 3D Detectors journal January 2015
Preliminary investigation of the dosimetric properties of ClearView™ dosimeter journal May 2017
AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology journal June 2001
High resolution gel-dosimetry by optical-CT and MR scanning journal July 2001
Determination of the initial beam parameters in Monte Carlo linac simulation: Monte Carlo Linac Simulation journal March 2006
Characterization of a new radiochromic three-dimensional dosimeter: Characterization of a new radiochromic 3D dosimeter journal April 2006
Basic investigations on the performance of a normoxic polymer gel with tetrakis-hydroxy-methyl-phosphonium chloride as an oxygen scavenger: Reproducibility, accuracy, stability, and dose rate dependence: Reproducibility, accuracy, stability, and dose rate dependence journal June 2006
MicroRT-Small animal conformal irradiator: MicroRT journal November 2007
Characterization of image quality and image-guidance performance of a preclinical microirradiator: Characterization of a preclinical microirradiator journal January 2011
Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques: Commissioning a small-field biological irradiator journal November 2011
Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71: Monitor unit calculations for photon and electron beams journal February 2014
Multi-institutional dosimetric and geometric commissioning of image-guided small animal irradiators: Multi-institutional commissioning of small animal irradiators journal March 2014
AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45 journal July 1994
Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 journal October 1998
AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams journal September 1999
Advanced Small Animal Conformal Radiation Therapy Device journal July 2016
Dose response characteristics and basic dose distribution data for a polymerization-based dosemeter gel evaluated using MR. journal January 2000
Specific issues in small animal dosimetry and irradiator calibration journal October 2011
Integrating Small Animal Irradiators withFunctional Imaging for Advanced Preclinical Radiotherapy Research journal February 2019
Dual-Modality X-Ray-Induced Radiation Acoustic and Ultrasound Imaging for Real-Time Monitoring of Radiotherapy journal May 2020