PSL-AC-CP-1102-004-03: Electrostatic Discharge Simulator Kit Model 930D and Associated Current-Viewing Resistor (V.04)
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
This document provides instructions for calibrating the Electro-Tech (ETS) electrostatic discharge (ESD) simulator, Model 930D. The calibration shall meet the ± 5% specification for resistance and capacitance as specified in MIL-STD-331C 2009 newer. A series of direct measurements of the Device Under Test (DUT) output at 25 kV using a calibrated LeCroy HDO 6104 oscilloscope (or equivalent 1 MΩ) input impedance storage oscilloscope) are recorded through the 500-pF capacitor, 500-Ω resistor, and a 1-Ω Current Viewing Resistor (CVR). The certified value of the DUT's 25 kV output is calculated using Ohms Law and the certified system resistance and capacitance described in this procedure. Read this document in its entirety before proceeding with the calibration.
- Research Organization:
- Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA)
- DOE Contract Number:
- AC04-94AL85000; NA0003525
- OSTI ID:
- 1762670
- Report Number(s):
- SAND--2019-7230R; PSL-AC-CP--1102-004-04; 676798
- Country of Publication:
- United States
- Language:
- English
Similar Records
Resistors applied to disconnect switches
Linear Induction Accelerator Cell Voltage Calibration Using Compact Electron Permanent Magnet Spectrometer
A least squares method for CVT calibration in a RLC capacitor discharge circuit.
Journal Article
·
Mon Jan 31 23:00:00 EST 1977
· Electr. World; (United States)
·
OSTI ID:7307020
Linear Induction Accelerator Cell Voltage Calibration Using Compact Electron Permanent Magnet Spectrometer
Technical Report
·
Tue Oct 08 00:00:00 EDT 2019
·
OSTI ID:1569726
A least squares method for CVT calibration in a RLC capacitor discharge circuit.
Technical Report
·
Fri Oct 31 23:00:00 EST 2003
·
OSTI ID:918301