Liquidus Temperature: Assessing Standard Glasses for Furnace Calibration

Technical Report ·
DOI:https://doi.org/10.2172/1615374· OSTI ID:1615374
Standard glasses are used to ensure accurate liquidus temperature measurements and furnace temperatures. The supply of the standard glasses used at Pacific Northwest National Laboratory (Standard Reference Material 773 [SRM-773]) has been depleted. Replacement standard glasses for furnace temperature verification and liquidus temperature measurement validation were fabricated and tested. Two glasses-ARG-1 and AmCm2-19-were fabricated in roughly 5 kg batches. The liquidus temperature of the glasses was measured and found to be consistent with those reported in literature.
Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1615374
Report Number(s):
PNNL--29312; EWG-RPT--022.Rev.0.0
Country of Publication:
United States
Language:
English

Similar Records

Solubility approach for modeling waste glass liquidus
Conference · Thu Jul 01 04:00:00 UTC 1999 · OSTI ID:20015763

The liquidus temperature of nuclear waste glasses: an international Round-Robin Study
Journal Article · Sat Dec 01 04:00:00 UTC 2012 · International Journal of Applied Glass Science, 2(4):321-333 · OSTI ID:1057825

Liquidus Temperature Data for DWPF Glass