Title: A second‐generation virtual‐pinhole PET device for enhancing contrast recovery and improving lesion detectability of a whole‐body PET/CT scanner

Journal Article · · Medical Physics
DOI:https://doi.org/10.1002/mp.13724· OSTI ID:1547606
 [1];  [2];  [1];  [3];  [3];  [3];  [2];  [1]
  1. Mallinckrodt Institute of Radiology, Washington University in St. Louis St. Louis MO 63110 USA
  2. Department of Electrical and Systems Engineering Washington University in St. Louis St. Louis MO 63130 USA
  3. Molecular Imaging Siemens Medical Solutions USA, Inc Knoxville TN 37932 USA

Purpose We have developed a second‐generation virtual‐pinhole (VP) positron emission tomography (PET) device that can position a flat‐panel PET detector around a patient's body using a robotic arm to enhance the contrast recovery coefficient (CRC) and detectability of lesions in any region‐of‐interest using a whole‐body PET/computed tomography (CT) scanner. Methods We constructed a flat‐panel VP‐PET device using 32 high‐resolution detectors, each containing a 4   4 MPPC array and 16   16 LYSO crystals of 1.0   1.0   3.0 mm 3 each. The flat‐panel detectors can be positioned around a patient's body anywhere in the imaging field‐of‐view (FOV) of a Siemens Biograph 40 PET/CT scanner by a robotic arm. New hardware, firmware and software have been developed to support the additional detector signals without compromising a scanner's native functions. We stepped a 22 Na point source across the axial FOV of the scanner to measure the sensitivity profile of the VP‐PET device. We also recorded the coincidence events measured by the scanner detectors and by the VP‐PET detectors when imaging phantoms of different sizes. To assess the improvement in the CRC of small lesions, we imaged an elliptical torso phantom measuring 316   228   162 mm 3 that contains spherical tumors with diameters ranging from 3.3 to 11.4 mm with and without the VP‐PET device. Images were reconstructed using a list mode Maximum‐Likelihood Estimation‐Maximization algorithm implemented on multiple graphics processing units (GPUs) to support the unconventional geometries enabled by a VP‐PET system. The mean and standard deviation of the CRC were calculated for tumors of different sizes. Monte Carlo simulation was also conducted to image clusters of lesions in a torso phantom using a PET/CT scanner alone or the same scanner equipped with VP‐PET devices. Receiver operating characteristic (ROC) curves were analyzed for three system configurations to evaluate the improvement in lesion detectability by the VP‐PET device over the native PET/CT scanner. Results The repeatability in positioning the flat‐panel detectors using a robotic arm is better than 0.15 mm in all three directions. Experimental results show that the average CRC of 3.3, 4.3, and 6.0 mm diameter tumors was 0.82%, 2.90%, and 5.25%, respectively, when measured by the native scanner. The corresponding CRC was 2.73%, 6.21% and 10.13% when imaged by the VP‐PET insert device with the flat‐panel detector under the torso phantom. These values may be further improved to 4.31%, 9.65% and 18.01% by a future dual‐panel VP‐PET insert device if DOI detectors are employed to triple its detector efficiency. Monte Carlo simulation results show that the tumor detectability can be improved by a VP‐PET device that has a single flat‐panel detector. The improvement is greater if the VP‐PET device employs a dual‐panel design. Conclusions We have developed a prototype flat‐panel VP‐PET device and integrated it with a clinical PET/CT scanner. It significantly enhances the contrast of lesions, especially for those that are borderline detectable by the native scanner, within regions‐of‐interest specified by users. Simulation demonstrated the enhancement in lesion detectability with the VP‐PET device. This technology may become a cost‐effective solution for organ‐specific imaging tasks.

Sponsoring Organization:
USDOE
Grant/Contract Number:
SC0005157
OSTI ID:
1547606
Journal Information:
Medical Physics, Journal Name: Medical Physics Vol. 46 Journal Issue: 9; ISSN 0094-2405
Publisher:
Wiley Blackwell (John Wiley & Sons)Copyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 6 works
Citation information provided by
Web of Science

References (38)

Targeted multi-pinhole SPECT journal November 2010
Feasibility study of a point‐of‐care positron emission tomography system with interactive imaging capability journal February 2019
Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo II journal September 2009
Initial Characterization of a Dedicated Breast PET/CT Scanner During Human Imaging journal August 2009
Study of a high-resolution PET system using a Silicon detector probe journal September 2014
Role of Positron Emission Tomography for the Monitoring of Response to Therapy in Breast Cancer journal January 2015
Combined PET/MR Imaging — Technology and Applications journal February 2010
GATE: A Geant4-Based Simulation Platform for PET and SPECT Integrating Movement and Time Management journal October 2003
A High-Resolution PET Demonstrator using a Silicon “Magnifying Glass” journal January 2012
Monitoring Tumor Response to Therapy with 18F-FLT PET journal May 2010
Fast List-Mode Reconstruction for Time-of-Flight PET Using Graphics Hardware journal February 2011
Adaptive Imaging for Lesion Detection Using a Zoom-in PET System journal January 2011
A compact high resolution flat panel PET detector based on the new 4-side buttable MPPC for biomedical applications
  • Wang, Qiang; Wen, Jie; Ravindranath, Bosky
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 794 https://doi.org/10.1016/j.nima.2015.04.071
journal September 2015
A combined time-of-flight and depth-of-interaction detector for total-body positron emission tomography: A combined time-of-flight and depth-of-interaction detector journal January 2016
Performance evaluation of a very high resolution small animal PET imager using silicon scatter detectors journal April 2007
Breast cancer detection using high-resolution breast PET compared to whole-body PET or PET/CT journal October 2013
Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small-Animal PET Scanner journal September 2008
Image reconstruction and system modeling techniques for virtual-pinhole PET insert systems journal April 2012
Comparison of quantitative diagnostic tests: Type I error, power, and sample size journal March 1987
Improving PET imaging for breast cancer using virtual pinhole PET half-ring insert journal September 2013
Fully 3D list-mode time-of-flight PET image reconstruction on GPUs using CUDA: Fully 3D list-mode time-of-flight PET on GPU-CUDA journal December 2011
Positron emission tomography with fluorine-18-deoxyglucose in the detection and staging of breast cancer journal June 1993
Radionuclide Methods and Instrumentation for Breast Cancer Detection and Diagnosis journal July 2013
Virtual-Pinhole PET journal February 2008
An evaluation of polygraphers' judgments: A review from a decision theoretic perspective. journal January 1982
A silicon PET probe
  • Studen, A.; Chesi, E.; Cindro, V.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 648 https://doi.org/10.1016/j.nima.2010.11.073
journal August 2011
Real time image reconstruction using GPUs for a surgical PET imaging probe system
  • Huh, S. S.; Rogers, W. L.
  • 2009 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC 2009), 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC) https://doi.org/10.1109/NSSMIC.2009.5402312
conference October 2009
Update on Time-of-Flight PET Imaging journal December 2014
Model-based normalization for iterative 3D PET image reconstruction journal July 2002
Total-Body PET: Maximizing Sensitivity to Create New Opportunities for Clinical Research and Patient Care journal September 2017
Clinical Performance of 2 Dedicated PET Scanners for Breast Imaging: Initial Evaluation journal August 2012
Imaging of pulmonary mass lesions with whole-body positron emission tomography and fluorodeoxyglucose journal July 1993
Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction journal June 2009
A simulation study for the design of a prototype insert for whole-body PET scanners journal June 2006
Simple charge division readouts for imaging scintillator arrays using a multi-channel PMT journal June 1996
Design of a very high-resolution small animal PET scanner using a silicon scatter detector insert journal July 2007
Theoretical analysis and simulation study of a high-resolution zoom-in PET system journal August 2009
Multi-resolution image reconstruction for a high-resolution small animal PET device conference January 2003