Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Automated 20 kpsi RPLC-MS and MS/MS with chromatographic peak capacities of 1000-1500 and capabilities in proteomics and metabolomics

Journal Article · · Analytical Chemistry, 77(10):3090-3100
DOI:https://doi.org/10.1021/ac0483062· OSTI ID:15016605

Proteomics analysis based-on liquid chromatography (LC), particularly reversed-phase LC (RPLC), is widely practiced; however, cutting-edge LC performance variations have generally not been adopted even though their benefits are well established. The two major reasons behind this general underutilization are: 1) uncertainties surrounding the extent of improvement (e.g., proteome coverage), and 2) the lack of availability of automated, robust, and convenient LC instrumentation. Here, we describe an automated format 20K psi gradient nanoscale LC system that was developed to provide improved separations and sensitivity for proteomics (and metabolomics) applications. The system includes on-line coupling of micro solid phase extraction for sample loading and allows emitters for electrospray ionization to be readily replaced. The system uses 40 to 200 cm X 50 µm i.d. fused silica capillaries packed with 1.4- to 3-µm porous C18-bonded silica particles to obtain chromatographic peak capacities of 1,000-1,500 for complex peptide and metabolite mixtures. This separation quality allowed high confidence identification of >12,000 different peptides from >2,000 distinct Shewanella oneidensis proteins (~ 40% of the proteins predicted for the S. oneidensis proteome) in a single 12-h ion trap tandem mass spectrometry (MS/MS) analysis. The reproducibility was >87% for proteins identified between replicates. The protein MS/MS identification rate average exceeded 10 proteins per minute, e.g., 1,207 proteins were identified in 120 min through assignment of 5,944 different peptides. For a human blood plasma sample that was not depleted of the most abundant proteins, 835 distinct proteins were identified with high confidence in a single 12-h run. A single run with accurate mass MS detected >5,000 different compounds from a metabolomics sample.

Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (US)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
15016605
Report Number(s):
PNNL-SA-43807; 400412000
Journal Information:
Analytical Chemistry, 77(10):3090-3100, Journal Name: Analytical Chemistry, 77(10):3090-3100 Journal Issue: 10 Vol. 77
Country of Publication:
United States
Language:
English

Similar Records

Ultrahigh-Throughput Proteomics Using Fast RPLC Separations with ESI-MS/MS
Journal Article · Sat Oct 15 00:00:00 EDT 2005 · Analytical Chemistry, 77(20):6692-6701 · OSTI ID:876924

Ultra-High-Efficiency Strong Cation Exchange LC/RPLC/MS/MS for High Dynamic Range Characterization of the Human Plasma Proteome
Journal Article · Sat Feb 14 23:00:00 EST 2004 · Analytical Chemistry, 76(4):1134-1144 · OSTI ID:15010465

Using size exclusion chromatography-RPLC and RPLC-CIEF as two-dimensional separation strategies for protein profiling
Journal Article · Sat Jul 01 00:00:00 EDT 2006 · Electrophoresis, 27(13):2722-2733 · OSTI ID:890108