Using Data Fusion to Characterize Breast Tissue

Conference ·
OSTI ID:15004026
New ultrasound data, obtained with a circular experimental scanner, are compared with data obtained with standard X-ray CT. Ultrasound data obtained by scanning fixed breast tissue were used to generate images of sound speed and reflectivity. The ultrasound images exhibit approximately 1 mm resolution and about 20 dB of dynamic range. All data were obtained in a circular geometry. X-ray CT scans were used to generate X-ray images corresponding to the same 'slices' obtained with the ultrasound scanner. The good match of sensitivity, resolution and angular coverage between the ultrasound and X-ray data makes possible a direct comparison of the three types of images. We present the results of such a comparison for an excised breast fixed in formalin. The results are presented visually using various types of data fusion. A general correspondence between the sound speed, reflectivity and X-ray morphologies is found. The degree to which data fusion can help characterize tissue is assessed by examining the quantitative correlations between the ultrasound and X-ray images.
Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
15004026
Report Number(s):
UCRL-JC-145038
Country of Publication:
United States
Language:
English

Similar Records

Characterizing Tissue with Acoustic Parameters Derived from Ultrasound Data
Conference · Tue Jan 22 23:00:00 EST 2002 · OSTI ID:15004024

Development of Ultrasound Tomography for Breast Imaging: Technical Assessment
Journal Article · Thu Sep 30 00:00:00 EDT 2004 · Medical Physics · OSTI ID:15016771

A Comparison of Ultrasound Tomography Methods in Circular Geometry
Conference · Wed Jan 23 23:00:00 EST 2002 · OSTI ID:15004025