Direct low-energy measurement of 37Ar and 127Xe in a radiotracer gas using low-background proportional counters
Journal Article
·
· Journal of Radioanalytical and Nuclear Chemistry
A radiotracer gas with a blend of 37Ar and 127Xe was created for a gas migration experiment and was characterized at Pacific Northwest National Laboratory using ultra-low-background proportional counters. This paper describes the direct low-energy measurement of 37Ar and 127Xe in a dual-isotope sample. Using this low-energy technique, the dual-isotope radiotracer gas was determined to have activity concentrations of 483 Bq/cc and 1,435 Bq/cc for 37Ar and 127Xe, respectively, and a ratio of 1:3 on the reference date of 7/11/2016.
- Research Organization:
- Pacific Northwest National Laboratory (PNNL), Richland, WA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC05-76RL01830
- OSTI ID:
- 1489257
- Report Number(s):
- PNNL-SA-133392
- Journal Information:
- Journal of Radioanalytical and Nuclear Chemistry, Journal Name: Journal of Radioanalytical and Nuclear Chemistry Journal Issue: 1 Vol. 318; ISSN 0236-5731
- Publisher:
- Springer
- Country of Publication:
- United States
- Language:
- English
Similar Records
Calibration of an Ultra-Low-Background Proportional Counter for Measuring 37Ar
Shielding concepts for low-background proportional counter arrays in surface laboratories
Initial Characterization of Unequal-Length, Low-Background Proportional Counters for Absolute Gas-Counting Applications
Conference
·
Thu Aug 08 00:00:00 EDT 2013
·
OSTI ID:1203934
Shielding concepts for low-background proportional counter arrays in surface laboratories
Journal Article
·
Sun Jan 31 23:00:00 EST 2016
· Applied Radiation and Isotopes
·
OSTI ID:1255357
Initial Characterization of Unequal-Length, Low-Background Proportional Counters for Absolute Gas-Counting Applications
Conference
·
Sat Aug 31 00:00:00 EDT 2013
·
OSTI ID:1178523