Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Zeptocalorimetry

Technical Report ·
DOI:https://doi.org/10.2172/1481628· OSTI ID:1481628
 [1];  [1];  [1]
  1. Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Here we present the development of a Zeptocalorimeter. The motivation for designing and implementing such a device is driven, ultimately, by its anticipated exceptional sensitivity (10-21 J/K, at 2K). Such a device would be highly valuable in detecting minute quantities of mass for threat detection, studying fundamental phonon physics, and detecting energetic dissipation events at the attojoule level. To date, the most sensitive calorimeter demonstrated in the literature at 2K has been developed by the Roukes group at Caltech, where they achieved an addendum heat capacity of 10-15 J/K with a 1/1000 sensitivity to external stimuli. To obtain such a low value of heat capacity requires a very small thermal mass, and thus, one of the greatest challenges in this project is the fabrication of this device, which requires numerous precision nanofabrication techniques. Furthermore, the heat capacity measurement of this device, as performed from room temperature to cryogenic temperatures, is equally challenging, as the transient signals used to determine the platform's thermal time constant require careful attention to the mitigation of feedthrough capacitance and delicate amplifier offsets. In this report we describe in detail the fabrication process flow for developing the calorimeter, including the layout and device design for obtaining a single lumped RC thermal resistance and capacitance, so that the device can be used for quantitative measurements of nanoscale materials with a suitable thermal link. The measurement method and experimental setup are also given, where we explain the heater and thermometer calibration methods, the thermal resistance measurements, the transient measurements, and lastly the cryogenic setup with intermediate frequency cabling and the thermal sinking of those lines.
Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
DOE Contract Number:
AC04-94AL85000; NA0003525
OSTI ID:
1481628
Report Number(s):
SAND--2018-11207; 669588
Country of Publication:
United States
Language:
English

Similar Records

Thermal transistor and thermometer based on Coulomb-coupled conductors
Journal Article · Tue Jul 23 20:00:00 EDT 2019 · Physical Review. B · OSTI ID:1612777

Sensitive power compensated scanning calorimeter for analysis of phase transformations in small samples
Journal Article · Wed Jun 15 00:00:00 EDT 2005 · Review of Scientific Instruments · OSTI ID:20722919

Related Subjects