Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Quantitative validation of a new coregistration algorithm

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.467857· OSTI ID:148087
; ; ;  [1];  [2]
  1. Columbia Univ., New York, NY (United States)
  2. Strichman Medical Equipment, Medfield, MA (United States)
A new coregistration software package, Neuro9OO Image Coregistration software, has been developed specifically for nuclear medicine. With this algorithm, the correlation coefficient is maximized between volumes generated from sets of transaxial slices. No localization markers or segmented surfaces are needed. The coregistration program was evaluated for translational and rotational registration accuracy. A Tc-99m HM-PAO split-dose study (0.53 mCi low dose, L, and 1.01 mCi high dose, H) was simulated with a Hoffman Brain Phantom with five fiducial markers. Translation error was determined by a shift in image centroid, and rotation error was determined by a simplified two-axis approach. Changes in registration accuracy were measured with respect to: (1) slice spacing, using the four different combinations LL, LH, HL, HH, (2) translational and rotational misalignment before coregistration, (3) changes in the step size of the iterative parameters. In all the cases the algorithm converged with only small difference in translation offset, 0 and 0. At 6 nun slice spacing, translational efforts ranged from 0.9 to 2.8 mm (system resolution at 100 mm, 6.8 mm). The converged parameters showed little sensitivity to count density. In addition the correlation coefficient increased with decreasing iterative step size, as expected. From these experiments, the authors found that this algorithm based on the maximization of the correlation coefficient between studies was an accurate way to coregister SPECT brain images.
OSTI ID:
148087
Report Number(s):
CONF-941061--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 42; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Automated coregistration and statistical analyses of SPECT brain images
Journal Article · Sun May 01 00:00:00 EDT 1994 · Journal of Nuclear Medicine · OSTI ID:198040

Automated coregistration of MTI spectral bands.
Conference · Mon Dec 31 23:00:00 EST 2001 · OSTI ID:976143

Correlation of SPECT images of a three-dimensional brain phantom using a surface fitting technique
Conference · Thu Oct 01 00:00:00 EDT 1992 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6706149