Title: Imaging of large-area microchannel plates using phosphor screens

Journal Article · · Journal of Instrumentation

A testing system was built, using UV illumination at 254 nm and a cathodoluminescent phosphor screen, to measure the gain uniformity of large area (20 x 20 cm(2)) microchannel plates. The phosphor screen image is captured with a wide-angle CMOS digital camera (4032 x 3024 pixels) and analyzed. First a reference MCP is illuminated and the generated phosphor screen image is captured with a CCD camera. In a second step, another image is captured with the MCP under test inserted between the reference MCP and the phosphor screen. The ratio of the two images is analyzed and provides a quantitative measure of the relative gain uniformity of the MCP under test.

Research Organization:
Argonne National Lab. (ANL), Argonne, IL (United States)
Sponsoring Organization:
USDOE Office of Science - Office of High Energy Physics; USDOE Office of Science (SC), Basic Energy Sciences (BES); National Natural Science Foundation of China (NNSFC)
DOE Contract Number:
AC02-06CH11357
OSTI ID:
1395999
Journal Information:
Journal of Instrumentation, Vol. 9, Issue 11; ISSN 1748-0221
Publisher:
Institute of Physics (IOP)
Country of Publication:
United States
Language:
English