Design and first plasma measurements of the ITER-ECE prototype radiometer
Journal Article
·
· Review of Scientific Instruments
- Univ. of Texas, Austin, TX (United States). Inst. for Fusion Studies; General Atomics
- Univ. of Texas, Austin, TX (United States). Inst. for Fusion Studies
- ITER Organization, Gandhinagar (India). Inst. for Plasma Research
- Virginia Diodes, Inc., Charlottesville, VA (United States)
On ITER, second harmonic optically thick electron cyclotron emission (ECE) in the range of 220-340 GHz will supply the electron temperature (Te). In order to investigate the requirements and capabilities prescribed for the ITER system, a prototype radiometer covering this frequency range has been developed by Virginia Diodes, Inc. The first plasma measurements with this instrument have been carried out on the DIII-D tokamak, with lab bench tests and measurements of third through fifth harmonic ECE from high Te plasmas. At DIII-D the instrument shares the transmission line of the Michelson interferometer and can simultaneously acquire data. In our comparison of the ECE radiation temperature from the absolutely calibrated Michelson and the prototype receiver we show that the ITER radiometer provides accurate measurements of the millimeter radiation across the instrument band.
- Research Organization:
- General Atomics, San Diego, CA (United States)
- Sponsoring Organization:
- USDOE
- Grant/Contract Number:
- FC02-04ER54698; FG02-97ER54415
- OSTI ID:
- 1371880
- Alternate ID(s):
- OSTI ID: 22596540
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 87; ISSN 0034-6748; ISSN RSINAK
- Publisher:
- American Institute of Physics (AIP)Copyright Statement
- Country of Publication:
- United States
- Language:
- English
Fusion Plasma Diagnostics with mm-Waves
|
book | August 2013 |
High resolution electron cyclotron emission temperature profile and fluctuation diagnostic for Alcator C-Mod
|
journal | January 1999 |
Control and dissipation of runaway electron beams created during rapid shutdown experiments in DIII-D
|
journal | July 2013 |
Electron cyclotron emission based q -profile measurement and concept for equilibrium reconstruction
|
journal | June 2019 |
Similar Records
Design and first plasma measurements of the ITER-ECE prototype radiometer
ECE RADIOMETER UPGRADE ON THE DIII-D TOKAMAK
Status of the design of the ITER ECE diagnostic
Journal Article
·
Mon Nov 14 23:00:00 EST 2016
· Review of Scientific Instruments
·
OSTI ID:22596540
ECE RADIOMETER UPGRADE ON THE DIII-D TOKAMAK
Conference
·
Thu Aug 01 00:00:00 EDT 2002
·
OSTI ID:813649
Status of the design of the ITER ECE diagnostic
Journal Article
·
Wed Mar 11 20:00:00 EDT 2015
· EPJ Web of Conferences
·
OSTI ID:1148889