Title: Comparison of analog and digital pulse-shape-discrimination systems

Journal Article · · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
 [1];  [1];  [1]
  1. Univ. of Michigan, Ann Arbor, MI (United States)

Pulse shape discrimination (PSD) performance of two optimized PSD systems (one digital and one analog) is compared in this work. One system uses digital charge integration, while the other system uses analog zero crossing. Measurements were conducted with each PSD system using the CAEN V1720 (250 MHz) data acquisition system. An organic-liquid scintillator, coupled to a photo-multiplier tube, was used to detect neutrons and gamma rays from a Cf-252 spontaneous-fission source. The PSD perfor-mance of both systems was optimized and quantified using a traditional figure-of-merit (FOM) approach. FOM's were found for three, 300 keVee light-output bins, spanning from 100 to 1000 keVee, and one larger bin from 100 to 1800 keVee. Digital PSD outperformed analog PSD in the lowest light-output bin by approximately 50%, and by 11% for the highest light-output bin.

Research Organization:
Univ. of Michigan, Ann Arbor, MI (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
NA0002534
OSTI ID:
1366592
Alternate ID(s):
OSTI ID: 1359648
Journal Information:
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 826; ISSN 0168-9002
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 27 works
Citation information provided by
Web of Science

References (5)

A fast module for pulse shape analysis journal December 1987
A versatile pulse shape discriminator for charged particle separation and its application to fast neutron time-of-flight spectroscopy journal November 1978
A new four channel pulse shape discriminator conference October 2007
Influence of sampling properties of fast-waveform digitizers on neutron−gamma-ray, pulse-shape discrimination for organic scintillation detectors
  • Flaska, Marek; Faisal, Muhammad; Wentzloff, David D.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 729 https://doi.org/10.1016/j.nima.2013.07.008
journal November 2013
A simple zero crossing pulse shape discrimination system journal May 1971

Similar Records

Pulse shape discrimination performance of stilbene coupled to low-noise silicon photomultipliers
Journal Article · Fri May 08 00:00:00 EDT 2015 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1366592

Digital front-end electronics for a tagged neutron inspection system
Journal Article · Fri Apr 19 00:00:00 EDT 2013 · AIP Conference Proceedings · OSTI ID:1366592

Application of deconvolution to recover frequency-domain multiplexed detector pulses
Journal Article · Thu Mar 21 00:00:00 EDT 2019 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1366592