Title: Comparison of film measurements and Monte Carlo simulations of dose delivered with very high‐energy electron beams in a polystyrene phantom

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4914371· OSTI ID:1348867

Purpose: To measure radiation dose in a water‐equivalent medium from very high‐energy electron (VHEE) beams and make comparisons to Monte Carlo (MC) simulation results. Methods: Dose in a polystyrene phantom delivered by an experimental VHEE beam line was measured with Gafchromic films for three 50 MeV and two 70 MeV Gaussian beams of 4.0–6.9 mm FWHM and compared to corresponding MC‐simulated dose distributions. MC dose in the polystyrene phantom was calculated with the EGSnrc/BEAMnrc and DOSXYZnrc codes based on the experimental setup. Additionally, the effect of 2% beam energy measurement uncertainty and possible non‐zero beam angular spread on MC dose distributions was evaluated. Results: MC simulated percentage depth dose (PDD) curves agreed with measurements within 4% for all beam sizes at both 50 and 70 MeV VHEE beams. Central axis PDD at 8 cm depth ranged from 14% to 19% for the 5.4–6.9 mm 50 MeV beams and it ranged from 14% to 18% for the 4.0–4.5 mm 70 MeV beams. MC simulated relative beam profiles of regularly shaped Gaussian beams evaluated at depths of 0.64 to 7.46 cm agreed with measurements to within 5%. A 2% beam energy uncertainty and 0.286° beam angular spread corresponded to a maximum 3.0% and 3.8% difference in depth dose curves of the 50 and 70 MeV electron beams, respectively. Absolute dose differences between MC simulations and film measurements of regularly shaped Gaussian beams were between 10% and 42%. Conclusions: The authors demonstrate that relative dose distributions for VHEE beams of 50–70 MeV can be measured with Gafchromic films and modeled with Monte Carlo simulations to an accuracy of 5%. The reported absolute dose differences likely caused by imperfect beam steering and subsequent charge loss revealed the importance of accurate VHEE beam control and diagnostics.

Sponsoring Organization:
USDOE
Grant/Contract Number:
DE‐AC02‐76SF00515; AC02-76SF00515
OSTI ID:
1348867
Journal Information:
Medical Physics, Journal Name: Medical Physics Vol. 42 Journal Issue: 4; ISSN 0094-2405
Publisher:
Wiley Blackwell (John Wiley & Sons)Copyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 30 works
Citation information provided by
Web of Science

References (18)

BEAM: A Monte Carlo code to simulate radiotherapy treatment units journal May 1995
PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter journal May 1995
Electron beam therapy is not inferior to superficial x-ray therapy in the treatment of skin carcinoma journal July 1995
Guidelines for evaluating and expressing the uncertainty of NIST measurement results report January 1994
Optimized treatment planning for prostate cancer comparing IMPT, VHEET and 15 MV IMXT journal June 2002
Benchmarking of the Dose Planning Method (DPM) Monte Carlo code using electron beams from a racetrack microtron journal May 2002
Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film: Absorbed-dose energy dependence of GAFCHROMIC film journal February 2010
Very High Energy Electrons (50 – 250 MeV) and Radiation Therapy journal April 2002
Precise radiochromic film dosimetry using a flat-bed document scanner: Radiochromic film dosimetry with flat-bed document scanner journal June 2005
Mycosis fungoides: radiation therapy journal December 2003
Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies: Energy dependence and dose response of Gafchromic EBT2 journal April 2010
Electron and Photon Beams from a 50 MeV Racetrack Microtron journal January 1980
Absolute response of Fuji imaging plate detectors to picosecond-electron bunches journal January 2010
150-250 MeV electron beams in radiation therapy journal June 2000
The Physics of Radiation Therapy journal March 2010
Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors: Dose rate dependence for different dosimeters an detectors journal April 2012
Experimental validation of the DPM Monte Carlo code using minimally scattered electron beams in heterogeneous media journal May 2002
Optimization of intensity-modulated very high energy (50–250 MeV) electron therapy journal April 2002