Preliminary Assessment of Microwave Readout Multiplexing Factor

Technical Report ·
DOI:https://doi.org/10.2172/1340978· OSTI ID:1340978
 [1];  [1];  [1];  [2];  [2];  [2];  [2];  [2];  [2]
  1. Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
  2. National Inst. of Standards and Technology (NIST), Boulder, CO (United States)

Ultra-high resolution microcalorimeter gamma spectroscopy is a new non-destructive assay technology for measurement of plutonium isotopic composition, with the potential to reduce total measurement uncertainty to a level competitive with destructive analysis methods [1-4]. Achieving this level of performance in practical applications requires not only the energy resolution now routinely achieved with transition-edge sensor microcalorimeter arrays (an order of magnitude better than for germanium detectors) but also high throughput. Microcalorimeter gamma spectrometers have not yet achieved detection efficiency and count rate capability that is comparable to germanium detectors, largely because of limits from existing readout technology. Microcalorimeter detectors must be operated at low temperature to achieve their exceptional energy resolution. Although the typical 100 mK operating temperatures can be achieved with reliable, cryogen-free systems, the cryogenic complexity and heat load from individual readout channels for large sensor arrays is prohibitive. Multiplexing is required for practical systems. The most mature multiplexing technology at present is time-division multiplexing (TDM) [3, 5-6]. In TDM, the sensor outputs are switched by applying bias current to one SQUID amplifier at a time. Transition-edge sensor (TES) microcalorimeter arrays as large as 256 pixels have been developed for X-ray and gamma-ray spectroscopy using TDM technology. Due to bandwidth limits and noise scaling, TDM is limited to a maximum multiplexing factor of approximately 32-40 sensors on one readout line [8]. Increasing the size of microcalorimeter arrays above the kilopixel scale, required to match the throughput of germanium detectors, requires the development of a new readout technology with a much higher multiplexing factor.

Research Organization:
Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
Sponsoring Organization:
USDOE Office of Nuclear Energy (NE)
DOE Contract Number:
AC52-06NA25396
OSTI ID:
1340978
Report Number(s):
LA-UR--16-27204
Country of Publication:
United States
Language:
English

Similar Records

14-pixel, multiplexed array of gamma-ray microcalorimeters with 47 eV energy resolution at 103 keV
Journal Article · Mon May 07 00:00:00 EDT 2007 · Applied Physics Letters · OSTI ID:20971919

A high resolution gamma-ray spectrometer based on superconducting microcalorimeters
Journal Article · Sat Sep 15 00:00:00 EDT 2012 · Review of Scientific Instruments · OSTI ID:22093728

Improved Nondestructive Isotopic Analysis with Practical Microcalorimeter Gamma Spectrometers
Journal Article · Tue Nov 30 23:00:00 EST 2021 · Journal of Nuclear Materials Management · OSTI ID:1999559