Accuracy evaluation of the LPTF optically pumped primary frequency standard
Conference
·
OSTI ID:125708
- Observatoire de Paris (France)
The optically pumped Cesium beam frequency standard under development in our laboratory was first evaluated in `93. The short term stability was {sigma}{sub y}({tau}) is 5.5 x 10{sup -13}, {tau}{sup -1/2}. The accuracy has been estimated to 1.1 x 10{sup -13}, mainly limited by the distributed phase shift. After modification of the cavity, a new evaluation is in progress and will be presented at the conference.
- OSTI ID:
- 125708
- Report Number(s):
- CONF-940603-; TRN: 95:005769-0075
- Resource Relation:
- Conference: CPEM `94: conference on precision electromagnetic measurements, Boulder, CO (United States), 27 Jun - 1 Jul 1994; Other Information: PBD: 1994; Related Information: Is Part Of 1994 Conference on precision electromagnetic measurements digest; DeWeese, E.; Bennett, G. [eds.]; PB: 577 p.
- Country of Publication:
- United States
- Language:
- English
Similar Records
A Cesium fountain frequency standard: Preliminary results
AN EVALUATION OF THE STABILITY AND ACCURACY OF A BREADBOARD MODEL OF A CESIUM GAS CELL ATOMIC FREQUENCY STANDARD. Technical Memorandum No. 747
Accuracy of Ultrasound-Based (BAT) Prostate-Repositioning: A Three-Dimensional On-Line Fiducial-Based Assessment With Cone-Beam Computed Tomography
Conference
·
Sat Dec 31 00:00:00 EST 1994
·
OSTI ID:125708
AN EVALUATION OF THE STABILITY AND ACCURACY OF A BREADBOARD MODEL OF A CESIUM GAS CELL ATOMIC FREQUENCY STANDARD. Technical Memorandum No. 747
Technical Report
·
Wed Jan 01 00:00:00 EST 1958
·
OSTI ID:125708
Accuracy of Ultrasound-Based (BAT) Prostate-Repositioning: A Three-Dimensional On-Line Fiducial-Based Assessment With Cone-Beam Computed Tomography
Journal Article
·
Sat Mar 15 00:00:00 EDT 2008
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:125708
+6 more