Direct tests of micro channel plates as the active element of a new shower maximum detector

Journal Article · · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
 [1];  [1];  [1];  [2];  [2];  [2];  [3]
  1. Fermi National Accelerator Lab. (FNAL), Batavia, IL (United States)
  2. California Inst. of Technology (CalTech), Pasadena, CA (United States)
  3. Univ. of Chicago, Chicago, IL (United States)
We continue the study of micro channel plates (MCP) as the active element of a shower maximum (SM) detector. We present below test beam results obtained with MCPs detecting directly secondary particles of an electromagnetic shower. The MCP efficiency to shower particles is close to 100%. Furthermore, the time resolution obtained for this new type of the SM detector is at the level of 40 ps.
Research Organization:
Fermi National Accelerator Laboratory (FNAL), Batavia, IL (United States)
Sponsoring Organization:
USDOE; USDOE Office of Science (SC), High Energy Physics (HEP) (SC-25)
Grant/Contract Number:
AC02-07CH11359; SC0011925
OSTI ID:
1235039
Alternate ID(s):
OSTI ID: 22461044
OSTI ID: 1247233
Report Number(s):
FERMILAB-PUB--15-296-PPD
Journal Information:
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Journal Name: Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment Journal Issue: C Vol. 795; ISSN 0168-9002
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English

Similar Records

Direct tests of a pixelated microchannel plate as the active element of a shower maximum detector
Journal Article · Fri May 06 20:00:00 EDT 2016 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1291341

Precision Timing with shower maximum detectors based on pixelated micro-channel plates
Journal Article · Sun Nov 26 19:00:00 EST 2017 · Journal of Physics. Conference Series · OSTI ID:1414908

Development of a New Fast Shower Maximum Detector Based on Microchannel Plates Photomultipliers (MCP-PMT) as an Active Element
Journal Article · Sun Sep 21 00:00:00 EDT 2014 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1296787