Design considerations for heated wells in gloveboxes.

Conference ·
OSTI ID:11738

Heated wells in gloveboxes have been used for many years by the Argonne National Laboratory Chemical Technology Division for nuclear-technology, waste-management, chemical-technology, and analytical-chemistry research. These wells allow experiments to be isolated from the main working volume of the glovebox. In addition, wells, when sealed, allow experiments to be conducted under pressurized or vacuum conditions. Until recently, typical maximum operational temperatures were about 500 C. However, more recent research is requiring operational temperatures approaching 900 C. These new requirements pose interesting design challenges that must be resolved. Some problem areas include temperature effects on material properties, maintaining a seal, cooling selected areas, and minimizing stresses. This paper discusses issues related to these design challenges and the ways in which these issues have been resolved.

Research Organization:
Argonne National Lab., IL (US)
Sponsoring Organization:
US Department of Energy (US)
DOE Contract Number:
W-31109-ENG-38
OSTI ID:
11738
Report Number(s):
ANL/CMT/CP-98693
Country of Publication:
United States
Language:
English

Similar Records

Redefining design criteria for Pu-238 gloveboxes
Conference · Wed Dec 30 23:00:00 EST 1998 · OSTI ID:319666

Minimizing Glovebox Glove Breaches: PART II.
Conference · Fri Dec 31 23:00:00 EST 2004 · OSTI ID:977947

Determining the Radiation Damage Effect on Glovebox Glove Material.
Conference · Fri Dec 31 23:00:00 EST 2004 · OSTI ID:977962