A novel assay method for the trace determination of Th and U in copper and lead using inductively coupled plasma mass spectrometry
Journal Article
·
· Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
This study describes a novel sample preparation and assay method developed in support of the MAJORANA DEMONSTRATOR experiment for the determination of thorium and uranium levels in copper and lead shielding components. Meticulously clean sample preparation methods combined with novel anion exchange separations for analyte pre-concentration and matrix removal were developed. Quantification was performed by inductively coupled plasma mass spectrometry. Detection limits of 0.0084 pg 232Th/g and 0.0106 pg 238U/g were determined for copper, while detection limits of 0.23 pg 232Th/g and 0.46 pg 238U/g were achieved for lead. These methods allow the Majorana Collaboration to accurately assay detector components and ensure that the experiment’s stringent radiopurity requirements are met.
- Research Organization:
- Pacific Northwest National Laboratory (PNNL), Richland, WA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC05-76RL01830
- OSTI ID:
- 1171869
- Report Number(s):
- PNNL-SA--104924; 830403000
- Journal Information:
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Journal Name: Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment Vol. 775; ISSN 0168-9002
- Publisher:
- Elsevier
- Country of Publication:
- United States
- Language:
- English
Similar Records
Mass spectrometric investigations into 3D printed parts to assess radiopurity as ultralow background materials for rare event physics detectors
Ultra-low radioactivity flexible printed cables
Journal Article
·
Mon Dec 05 19:00:00 EST 2022
· Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
·
OSTI ID:1967945
Ultra-low radioactivity flexible printed cables
Journal Article
·
Mon Sep 18 20:00:00 EDT 2023
· EPJ Techniques and Instrumentation
·
OSTI ID:2001029
Related Subjects
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
CONCENTRATION RATIO
COPPER
ICP MASS SPECTROSCOPY
ION EXCHANGE
Impurities
LEAD
REMOVAL
Radiation Detectors
SAMPLE PREPARATION
SENSITIVITY
SHIELDING
THORIUM 232
Trace Amounts
URANIUM 238
Ultratrace
Uranium
Thorium
Radiopurity Assay
ICP-MS
Majorana
CONCENTRATION RATIO
COPPER
ICP MASS SPECTROSCOPY
ION EXCHANGE
Impurities
LEAD
REMOVAL
Radiation Detectors
SAMPLE PREPARATION
SENSITIVITY
SHIELDING
THORIUM 232
Trace Amounts
URANIUM 238
Ultratrace
Uranium
Thorium
Radiopurity Assay
ICP-MS
Majorana