Test Report for NG Sensors GTX-1000
- Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
This report describes initial testing of the NG Sensor GTX-1000 natural gas monitoring system. This testing showed that the retention time, peak area stability and heating value repeatability of the GTX-1000 were promising for natural gas measurements in the field or at the well head. The repeatability can be less than 0.25% for LHV and HHV for the Airgas standard tested in this report, which is very promising for a first generation prototype. Ultimately this system should be capable of 0.1% repeatability in heating value at significant size and power reductions compared with competing systems.
- Research Organization:
- Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA)
- DOE Contract Number:
- AC04-94AL85000
- OSTI ID:
- 1170405
- Report Number(s):
- SAND2015--0878; 565603
- Country of Publication:
- United States
- Language:
- English