Title: Controlling Low-Rate Signal Path Microdischarge for an Ultra-Low-Background Proportional Counter

Journal Article · · Journal of Radioanalytical and Nuclear Chemistry, 296(2):753-758

ABSTRACT Pacific Northwest National Laboratory (PNNL) has developed an ultra-low-background proportional counter (ULBPC) made of high purity copper. These detectors are part of an ultra-low-background counting system (ULBCS) in the newly constructed shallow underground laboratory at PNNL (at a depth of ~30 meters water-equivalent). To control backgrounds, the current preamplifier electronics are located outside the ULBCS shielding. Thus the signal from the detector travels through ~1 meter of cable and is potentially susceptible to high voltage microdischarge and other sources of electronic noise. Based on initial successful tests, commercial cables and connectors were used for this critical signal path. Subsequent testing across different batches of commercial cables and connectors, however, showed unwanted (but still low) rates of microdischarge noise. To control this noise source, two approaches were pursued: first, to carefully validate cables, connectors, and other commercial components in this critical signal path, making modifications where necessary; second, to develop a custom low-noise, low-background preamplifier that can be integrated with the ULBPC and thus remove most commercial components from the critical signal path. This integrated preamplifier approach is based on the Amptek A250 low-noise charge-integrating preamplifier module. The initial microdischarge signals observed are presented and characterized according to the suspected source. Each of the approaches for mitigation is described, and the results from both are compared with each other and with the original performance seen with commercial cables and connectors.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
1117067
Report Number(s):
PNNL-SA-86931; NN2001000
Journal Information:
Journal of Radioanalytical and Nuclear Chemistry, 296(2):753-758, Vol. 296, Issue 2; ISSN 0236--5731
Country of Publication:
United States
Language:
English

Similar Records

The Design, Construction, and Initial Characterization of an Ultra-Low-Background Gas-Proportional Counting System
Journal Article · Wed May 01 00:00:00 EDT 2013 · Journal of Radioanalytical and Nuclear Chemistry, 296(2):915-921 · OSTI ID:1117067

Calibration of an Ultra-Low-Background Proportional Counter for Measuring 37Ar
Conference · Thu Aug 08 00:00:00 EDT 2013 · OSTI ID:1117067

Digital Pulse-Shape Discrimination Applied to an Ultra-Low-Background Gas-Proportional Counting System: First Results
Journal Article · Wed May 01 00:00:00 EDT 2013 · Journal of Radioanalytical and Nuclear Chemistry, 296(2):823-827 · OSTI ID:1117067