A Design for a Compact Time-of-Flight Mass Spectrometer
The design of a prototype, compact time-of-flight (TOF) mass spectrometer (MS) is described. The system primarily consists of an ion acceleration/focusing/steering assembly (AFSA), an 8 cm field-free region, a 4 cm, dual-stage reflectron and a miniature microchannel plate detector. Consequently, the resulting flight length of the system is 12 cm. The system has been designed with the capability to sample directly from atmosphere at ambient pressures. This is accomplished through the use of an electrodynamic ion funnel, housed in an intermediate-vacuum chamber that is coupled to the inlet of the TOF chamber. TOF spectra were obtained using noble gases (Ar, Kr and Xe) as test chemicals. These measured flight times were used to probe the performance of the instrument. A temporal resolution (tflight/Δt) of approximately 125, acquired using 129Xe+, has been measured for the system.
- Research Organization:
- Nevada Test Site (NTS), Mercury, NV (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA)
- DOE Contract Number:
- DE-AC52-06NA25946
- OSTI ID:
- 1056101
- Report Number(s):
- DOE/NV/25946-1622
- Journal Information:
- Review of Scientific Instruments, Vol. 83, Issue 10
- Country of Publication:
- United States
- Language:
- English
Similar Records
Ion Mobility Spectrometry – Mass Spectrometry Performance Using Electrodynamic Ion Funnels and Elevated Drift Gas Pressures
Development of a hand-portable TOF mass spectrometer system for nonproliferation applications