Calibrating Accelerometers Using an Electromagnetic Launcher

Conference · · I2MTC Conference Proceedings
OSTI ID:1042647

A Pulse Forming Network (PFN), Helical Electromagnetic Launcher (HEML), Command Module (CM), and Calibration Table (CT) were built and evaluated for the combined ability to calibrate an accelerometer. The PFN has a maximum stored energy of 19.25 kJ bank and is fired by a silicon controlled rectifier (SCR), with appropriate safety precautions. The HEML is constructed out of G-10 fiberglass and is designed to accelerate 600 grams to 10 meters per second. The CM is microcontroller based running Arduino Software. The CM has a keypad input and 7 segment outputs of the bank voltage and desired voltage. After entering a desired bank voltage, the CM controls the charge of the PFN. When the two voltages are equal it allows the fire button to send a pulse to the SCR to fire the PFN and in turn, the HEML. The HEML projectile's tip hits a target that is held by the CT. The CT consists of a table to hold the PFN and HEML, a vacuum chuck, air bearing, velocity meter and catch pot. The Target is held with the vacuum chuck awaiting impact. After impact, the air bearing allows the target to fall freely for the velocity meter to get an accurate reading. A known acceleration is determined from the known change in velocity of the target. Thus, if an accelerometer was attached to the target, the measured value can be compared to the known value.

Research Organization:
Kansas City Plant (KCP), Kansas City, MO (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
DOE Contract Number:
AC04-01AL66850
OSTI ID:
1042647
Report Number(s):
KCP-613-9109
Journal Information:
I2MTC Conference Proceedings, Journal Name: I2MTC Conference Proceedings
Country of Publication:
United States
Language:
English

Similar Records

Measurement Uncertainty Analysis of an Accelerometer Calibration Using a POC Electromagnetic Launcher
Journal Article · Tue Jun 12 00:00:00 EDT 2012 · IEEE Transactions on Instrumentation and measurement · OSTI ID:1134044

Precision shock accelerometer calibrator
Conference · Wed Dec 31 23:00:00 EST 1975 · OSTI ID:7365299

Method and apparatus for determining a normal combustion characteristic for an internal combustion engine from an accelerometer signal
Patent · Mon Dec 12 23:00:00 EST 2011 · OSTI ID:1531708