A phase-shifting diffraction interferometer (PSDI) has been integrated into an adaptive optics (AO) system developed by LLNL for use on the three meter Shane telescope at Lick Observatory. The interferometer is an all fiber optic design, which is extremely compact. It is useful for calibrating the control sensors, measuring the aberrations of the entire AO optical train, and measuring the influence functions of the individual actuators on the deformable mirror. The PSDI is particularly well suited for this application because it measures converging, quasi-spherical wavefronts, such as are produced by an AO imaging system. Thus, a PSDI can be used to measure the aberrations of the entire AO system, in-situ and without errors introduced by auxiliary optics. This provides an extremely accurate measurement ({approximately} 5 nm RMS) of the optical properties of the AO system.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE Office of Defense Programs (DP) (US)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
10421
Report Number(s):
UCRL-JC-134805; YN0100000; 97-ERD-037; YN0100000; 97-ERD-037; TRN: AH200126%%235
Resource Relation:
Conference: Society of Photo-Optical Instrumentation Engineers, SPIE's 44th Annual Meeting, Denver, CO (US), 07/18/1999--07/23/1999; Other Information: PBD: 23 Jun 1999