D-Zero Assembly Hall ODH Analysis

Technical Report ·
DOI:https://doi.org/10.2172/1033268· OSTI ID:1033268
 [1]
  1. Fermi National Accelerator Laboratory (FNAL), Batavia, IL (United States)

The ODH analysis presented here covers the high bay and assembly hall docking area of the for the D-Zero detector. It includes the STand Alone helium Refrigerator (STAR) in the building. It also includes the D-Zero detector and it's associated cryogenic and gas systems. An ODH analysis is presented which shows that the D-Zero assembly building high bay including the detector docking area is ODH class O. Probabilities, leak rates, and fatality factors are generated for all items that are sources of inert gas. The scope of analysis included the calorimeter and gas components on the detector, the helium refrigerator/liquifier components, and the future solenoid and visible light photon counter cryogenics that will be added to the D-Zero detector. The analysis demonstrates that the calorimeter and helium refrigerator systems pose no ODH hazard to personnel.

Research Organization:
Fermi National Accelerator Laboratory (FNAL), Batavia, IL (United States)
Sponsoring Organization:
USDOE Office of Science (SC)
DOE Contract Number:
AC02-07CH11359
OSTI ID:
1033268
Report Number(s):
FERMILAB-D0--EN-463
Country of Publication:
United States
Language:
English

Similar Records

Commissioning Test Results for D-Zero's Helium Refrigerator
Technical Report · Sun Jun 29 20:00:00 EDT 1997 · OSTI ID:1033274

Oxygen deficiency hazard (ODH) analysis for pressure and cryogenic systems facilities
Conference · Tue Oct 31 23:00:00 EST 1995 · OSTI ID:122569

D0 Detector Assemble Hall Platform Oxygen Deficiency Hazard Analysis
Technical Report · Mon Jan 28 23:00:00 EST 1991 · OSTI ID:1031837