Shortwave Spectroradiometer (SWS) Handbook
The Shortwave Spectroradiometer (SWS), measures the absolute visible and near infrared spectralradiance (units of watts per meter square per nanometer per steradian) of the zenith directly above the instrument. The SWS is a moderate resolution sensor comprised of two Zeiss spectrometers (MMS 1 NIR enhanced and NIR-PGS 2.2) for visible and near-infrared detection in the wavelength range 350 – 2170 nm. The sampling frequency is 1 Hz. The spectral resolution is 8 nm for the MMS 1 NIR and 12 nm for the NIR-PGS 2.2. The light collector is a narrow field of view (1.4°) collimator at the front end of a high-grade custom-made fiber optic bundle. The SWS does daily daytime measurements. The SWS is located in a darkroom, constructed by Southern Great Plains (SGP) site personnel within the optical trailer, to permit calibrations to be conducted without the necessity of moving the instrument to a different location. Calibrations are performed at regularly scheduled times using the ARM 12” integrating sphere.
- Research Organization:
- DOE Office of Science Atmospheric Radiation Measurement (ARM) Program (United States)
- Sponsoring Organization:
- USDOE EM Office of Program and Site Support (EM-50)
- OSTI ID:
- 1020259
- Report Number(s):
- DOE/SC-ARM/TR-062
- Country of Publication:
- United States
- Language:
- English
Similar Records
Final Technical Report
ARM: Shortwave spectroradiometer (SWS): spectral data