Performance checks with the Alpha Sentry CAM

Conference ·
OSTI ID:10171513
Before a CAM is put into service, it must be calibrated. The flow meter and detector must be calibrated with an external flow meter to provide accurate flow data, and the detector must be calibrated to produce accurate DPM data. Both flow and DPM data enter into the calculation of the Derived Air Concentration exposure (DAC-hr) by the CAM software. The focus of this report is on methods for checking that the DAC-hr alarm functionality has been properly calibrated and available in installed CAM instruments. The process begins with detector calibration. In order to calibrate the detector, the Alpha Sentry CAM is placed in an off-line calibration mode and detector efficiency calibration is selected. The user is prompted to enter the calibration source DPM and place the source in the CAM head. Upon latching the filter door with the source in place, a count is automatically initiated and completed. From the count data and the user entered DPM data, an efficiency is determined in the Alpha Sentry Manager (ASM) and stored in non-volatile memory in the CAM head electronics. This source is typically a plated {sup 239}Pu or {sup 241}Am source diffused into a stainless steel planchet.