Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Hydrogenated amorphous silicon (a-Si:H) based gamma camera: Monte Carlo simulations

Conference ·
OSTI ID:10125682
A new gamma camera using a-Si:H photodetectors has been designed for the imaging of heart and other small organs. In this new design the photomultiplier tubes and the position sensing circuitry are replaced by 2-D array of a-Si:H p-i-n pixel photodetectors and readout circuitry which are built on a substrate. Without the photomultiplier tubes this camera is light weight, hence can be made portable. To predict the characteristics and the performance of this new gamma camera we did Monte Carlo simulations. In the simulations 128 {times} 128 imaging array of various pixel sizes were used. {sup 99m}Tc (140keV) and {sup 201}Tl(70keV) were used as radiation sources. From the simulations we could obtain the resolution of the camera and the overall system, and the blurring effects due to scattering in the phantom. Using the Wiener filter for image processing, restoration of the blurred image could be achieved. Simulation results of a-Si:H based gamma camera were compared with those of a conventional gamma camera.
Research Organization:
Lawrence Berkeley Lab., CA (United States)
Sponsoring Organization:
USDOE, Washington, DC (United States)
DOE Contract Number:
AC03-76SF00098
OSTI ID:
10125682
Report Number(s):
LBL--35050; CONF-940254--1; ON: DE94007042
Country of Publication:
United States
Language:
English