Calibration of a Flat Field Soft X-ray Grating Spectrometer for Laser Produced Plasmas

Conference ·
DOI:https://doi.org/10.1063/1.3495790· OSTI ID:1009655

We have calibrated the x ray response of a variable line spaced grating spectrometer, known as the VSG, at the Fusion and Astrophysics Data and Diagnostic Calibration Facility at the Lawrence Livermore National Laboratory (LLNL). The VSG has been developed to diagnose laser produced plasmas, such as those created at the Jupiter Laser Facility and the National Ignition Facility at LLNL, and at both the Omega and Omega EP lasers at University of Rochester's Laboratory for Laser Energetics. The bandwidth of the VSG spans the range from {approx} 6 to 60 {angstrom}. The calibration results present here include the VSG's dispersion and quantum efficiency. The dispersion is determined by measuring the x rays emitted from hydrogen-like and helium-like ions of carbon, nitrogen, oxygen, neon, and aluminum. The quantum efficiency is calibrated to an accuracy of 30% or better by normalizing the x ray intensities recorded by the VSG to those simultaneously recorded by an x ray microcalorimeter spectrometer.

Research Organization:
Lawrence Livermore National Laboratory (LLNL), Livermore, CA
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
1009655
Report Number(s):
LLNL-CONF-431711
Country of Publication:
United States
Language:
English

References (13)

Performance of the EBIT calorimeter spectrometer journal October 2008
Development of a time-resolved soft x-ray spectrometer for laser produced plasma experiments journal October 2010
OZSPEC-2: An improved broadband high-resolution elliptical crystal x-ray spectrometer for high-energy density physics experiments (invited) journal October 2008
Milestones in EBIT spectroscopy and why it almost did not work journal January 2008
The Suzaku High Resolution X-Ray Spectrometer journal January 2007
Rapid, absolute calibration of x-ray filters employed by laser-produced plasma diagnostics journal October 2008
A “brief” history of spectroscopy on EBIT journal January 2008
The National Ignition Facility: Ushering in a new age for high energy density science journal April 2009
Testing QED and atomic-nuclear interactions with high- Z ions journal March 2010
Development of a thermal X-radiation source using “hot” hohlraums journal May 2007
Demonstration of transient gain x-ray lasers near 20??nm for nickellike yttrium, zirconium, niobium, and molybdenum journal January 1999
Initial performance results of the OMEGA laser system journal January 1997
Self‐calibration of a thinned, backside illuminated charge coupled devices in the soft x‐ray region journal January 1995