Title: Multivariate statistical analysis of time-of-flight secondary ion mass spectrometry images : looking beyond the obvious.

Conference ·
OSTI ID:1005074

Analytical instrumentation such as time-of-flight secondary ion mass spectrometry (ToF-SIMS) provides a tremendous quantity of data since an entire mass spectrum is saved at each pixel in an ion image. The analyst often selects only a few species for detailed analysis; the majority of the data are not utilized. Researchers at Sandia National Laboratory (SNL) have developed a powerful multivariate statistical analysis (MVSA) toolkit named AXSIA (Automated eXpert Spectrum Image Analysis) that looks for trends in complete datasets (e.g., analyzes the entire mass spectrum at each pixel). A unique feature of the AXSIA toolkit is the generation of intuitive results (e.g., negative peaks are not allowed in the spectral response). The robust statistical process is able to unambiguously identify all of the spectral features uniquely associated with each distinct component throughout the dataset. General Electric and Sandia used AXSIA to analyze raw data files generated on an Ion Tof IV ToF-SIMS instrument. Here, we will show that the MVSA toolkit identified metallic contaminants within a defect in a polymer sample. These metallic contaminants were not identifiable using standard data analysis protocol.

Research Organization:
Sandia National Laboratories (SNL), Albuquerque, NM, and Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC04-94AL85000
OSTI ID:
1005074
Report Number(s):
SAND2003-3229C; TRN: US201105%%28
Resource Relation:
Conference: Proposed for presentation at the Fourteenth International Conference on Secondary Ion Mass Spectrometry (SIMS XIV) held September 14-19, 2003 in San Diego, CA.
Country of Publication:
United States
Language:
English