High Bias of DCAP and Aa Analyses of Hg(I)
This brief, but definitive, study indicates that the relatively new spectrometric techniques, DCAP and AA (Atomic Absorption), give high values for mercury when employed to analyze solutions that contain Hg(I) per se or that contain soluble mercury which is or has been subject to reducing conditions.