Title: Evaluation of thermoluminescent dosimeters (TLDs) of two different designs for beta particle and low energy photon dosimetry

Technical Report ·
DOI:https://doi.org/10.2172/6567527· OSTI ID:6567527
;  [1];  [2]
  1. Oak Ridge National Lab., TN (United States)
  2. Tennessee Univ., Knoxville, TN (United States). Dept. of Nuclear Engineering

The purpose of this research was to compare the response of the thermoluminescent dosimeters (TLDs) currently used at Martin Marietta Energy Systems with that of a newly designed TLD which is now under consideration. The new TLD consists of a thicker chip in element No. 3 and a thinner, more opaque mylar filter over this same element. The goals were to determine if the new (proposed) TLD would perform significantly different from the old (current) TLD when measuring dose equivalents from beta particles and low energy photons (x-rays) and to find out if changing from the old to the new design would require that modifications be made to the calibration factors currently used in the dose calculation algorithm. To accomplish these objectives, tests were performed using both types of dosimeters placed under identical conditions. Their responses were then compared and analyzed. It was concluded that the new TLDs would allow for the detection of lower levels of radiation and give more precise readouts at very low (i.e., background) levels in addition to the obvious advantages of the new design, such as, a thicker, more durable chip and a more opaque mylar. It was also determined that no adjustments to the calibration factors would be necessary in switching to the new design.

Research Organization:
Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE; USDOE, Washington, DC (United States)
DOE Contract Number:
AC05-84OR21400
OSTI ID:
6567527
Report Number(s):
ORNL-6737; ON: DE93012506
Country of Publication:
United States
Language:
English

Similar Records

Design of an advanced TLD-based fixed nuclear accident dosimeter
Technical Report · Sun Aug 01 00:00:00 EDT 1993 · OSTI ID:6567527

Determination of Radiation Energy Response for Thermoluminescent Dosimeter TLD-100: Determination of Organ Dose in Diagnostic Radiology
Journal Article · Sun Apr 19 00:00:00 EDT 2009 · AIP Conference Proceedings · OSTI ID:6567527

SIXTH ERDA WORKSHOP ON PERSONNEL NEUTRON DOSIMETRY
Technical Report · Mon Jul 11 00:00:00 EDT 1977 · OSTI ID:6567527