Title: Tissue-equivalent torso phantom for calibration of transuranic-nuclide counting facilities

Conference ·
OSTI ID:5796207

Several tissue-equivalent human-torso phantoms have been constructed for the calibration of counting systems used for in-vivo measurement of transuranic radionuclides. The phantoms contain a simulated human rib cage (in some cases, real bone) and removable model organs, and they include tissue-equivalent chest plates that can be placed over the torso to simulate people with a wide range of statures. The organs included are the lungs, liver, and tracheobronchial lymph nodes. Polyurethane with varying concentrations of added calcium was used to simulate the linear photon-attenuation properties of various human tissues, including lean muscle, adipose-muscle mixtures, cartilage, and bone. Foamed polyurethane was used to simulate lung tissue. Organs have been loaded with highly pure /sup 238/Pu, /sup 239/Pu, /sup 241/Am, and other radionuclides of interest. The validity of the phantom as a calibration standard has been checked in separate intercomparison studies using human subjects whose lungs contained a plutonium simulant. The resulting phantom calibration factors generally compared to within +-20% of the average calibration factors obtained for the human subjects.

Research Organization:
Lawrence Livermore National Lab., CA (USA)
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
5796207
Report Number(s):
UCRL-93776; CONF-860146-8; ON: DE86009693
Resource Relation:
Conference: Radiobioassay and internal dosimetry workshop, Albuquerque, NM, USA, 20 Jan 1986; Other Information: Portions of this document are illegible in microfiche products
Country of Publication:
United States
Language:
English