A fully automated precise electrical resistance measurement system
- Inter-University Consortium for DAE Facilities, University Campus, Indore 452001 (India)
A fully automated precise electrical resistance measurement system for more than one sample has been constructed. Conventional four-probe measurements with van der Pauw and Montgomery configurations are possible with this system. Resistance measurements in the range of a few {mu}{Omega} to a few G{Omega} are possible for six samples at a time from room temperature down to liquid-helium or liquid-nitrogen temperatures with a temperature control accuracy of better than 10 mK. The design features of the system with special reference to the low-noise switching methods of currents and voltages are described in detail. Precision of the results thus obtained using this system are highlighted for a few superconducting and semiconducting samples. {copyright} {ital 1996 American Institute of Physics.}
- OSTI ID:
- 286879
- Journal Information:
- Review of Scientific Instruments, Vol. 67, Issue 8; Other Information: PBD: Aug 1996
- Country of Publication:
- United States