Title: Application of A150 plastic-equivalent gas in A150 plastic ionization chambers for Co-60. gamma. rays and 14. 8-MeV neutrons

Journal Article · · Med. Phys.; (United States)
OSTI ID:5497639

An ion-chamber gas mixture having an atomic composition similar to that of A150 plastic has been tested experimentally. This ''A150 gas'', air, and methane-based Te gas were each flowed through A150 plastic-walled ion chambers of various sizes, and irradiated with /sup 60/CO ..gamma.. rays and 14.8-MeV neutrons, separately administered. Observed ionization was measured and corrected for recombination losses. Best values for W-bar/sub a150/ , averaged over all chamber sizes, were 26.2 eV/ion pair for /sup 60/CO ..gamma.. rays and 28.2 eV/ion pair for 14.8-MeV neutrons.

Research Organization:
Medical Physics Department, Deparment of Radiology, University of Wisconsin 53706
OSTI ID:
5497639
Journal Information:
Med. Phys.; (United States), Vol. 9:3
Country of Publication:
United States
Language:
English