Title: Characteristics of various tissue and A150-plastic equivalent gas mixtures for microdosimetric applications

Technical Report ·
OSTI ID:5018035

The relative gas gain, alpha particle resolution and gain stability of two A150-plastic equivalent gas mixtures (A150 Mix-1 and A150 Mix-5) as well as propane and methane based tissue equivalent mixtures were measured with a spherical A150-plastic-walled proportional counter. The event-size distribution in a 14.8-MeV neutron field was determined for each gas mixture. Absolute tissue doses deduced from these measurements were in substantial (1%) agreement with doses determined by a TE-plastic ionization chamber and miniature GM counter.

Research Organization:
Wisconsin Univ., Madison (USA). Dept. of Medical Physics
DOE Contract Number:
AC02-76EV01105
OSTI ID:
5018035
Report Number(s):
DOE/EV/01105-292; ON: DE82019973; TRN: 82-019695
Country of Publication:
United States
Language:
English