Title: Bowtie filters for dedicated breast CT: Theory and computational implementation

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4908002· OSTI ID:1401377

Purpose: To design bowtie filters with improved properties for dedicated breast CT to improve image quality and reduce dose to the patient. Methods: The authors present three different bowtie filters designed for a cylindrical 14‐cm diameter phantom with a uniform composition of 40/60 breast tissue, which vary in their design objectives and performance improvements. Bowtie design #1 is based on single material spectral matching and produces nearly uniform spectral shape for radiation incident upon the detector. Bowtie design #2 uses the idea of basis material decomposition to produce the same spectral shape and intensity at the detector, using two different materials. Bowtie design #3 eliminates the beam hardening effect in the reconstructed image by adjusting the bowtie filter thickness so that the effective attenuation coefficient for every ray is the same. All three designs are obtained using analytical computational methods and linear attenuation coefficients. Thus, the designs do not take into account the effects of scatter. The authors considered this to be a reasonable approach to the filter design problem since the use of Monte Carlo methods would have been computationally intensive. The filter profiles for a cone‐angle of 0° were used for the entire length of each filter because the differences between those profiles and the correct cone‐beam profiles for the cone angles in our system are very small, and the constant profiles allowed construction of the filters with the facilities available to us. For evaluation of the filters, we used Monte Carlo simulation techniques and the full cone‐beam geometry. Images were generated with and without each bowtie filter to analyze the effect on dose distribution, noise uniformity, and contrast‐to‐noise ratio (CNR) homogeneity. Line profiles through the reconstructed images generated from the simulated projection images were also used as validation for the filter designs. Results: Examples of the three designs are presented. Initial verification of performance of the designs was done using analytical computations of HVL, intensity, and effective attenuation coefficient behind the phantom as a function of fan‐angle with a cone‐angle of 0°. The performance of the designs depends only weakly on incident spectrum and tissue composition. For all designs, the dynamic range requirement on the detector was reduced compared to the no‐bowtie‐filter case. Further verification of the filter designs was achieved through analysis of reconstructed images from simulations. Simulation data also showed that the use of our bowtie filters can reduce peripheral dose to the breast by 61% and provide uniform noise and CNR distributions. The bowtie filter design concepts validated in this work were then used to create a computational realization of a 3D anthropomorphic bowtie filter capable of achieving a constant effective attenuation coefficient behind the entire field‐of‐view of an anthropomorphic breast phantom. Conclusions: Three different bowtie filter designs that vary in performance improvements were described and evaluated using computational and simulation techniques. Results indicate that the designs are robust against variations in breast diameter, breast composition, and tube voltage, and that the use of these filters can reduce patient dose and improve image quality compared to the no‐bowtie‐filter case.

Sponsoring Organization:
USDOE
OSTI ID:
1401377
Journal Information:
Medical Physics, Journal Name: Medical Physics Vol. 42 Journal Issue: 3; ISSN 0094-2405
Publisher:
Wiley Blackwell (John Wiley & Sons)Copyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 10 works
Citation information provided by
Web of Science

References (30)

A method for comparing beam-hardening filter materials for diagnostic radiology: Comparison of beam-hardening filter materials for diagnostic radiology journal July 1988
Computational Methods for the Design of Test Objects and Tissue Substitutes for Radiologic Applications journal September 1993
Cone-beam volume CT breast imaging: Feasibility study journal April 2002
Dedicated Cone-Beam Breast CT: Feasibility Study with Surgical Mastectomy Specimens journal December 2007
Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner: Scatter properties in cone-beam breast CT scanner journal August 2005
The myth of the 50-50 breast: Myth of 50-50 breast journal November 2009
Energy-resolved computed tomography: first experimental results journal September 2008
Practical cone-beam algorithm journal January 1984
Compensators for dose and scatter management in cone-beam computed tomography: Compensators for cone-beam CT journal June 2007
Achieving uniform noise in direct fan-beam CT reconstruction through bowtie filter design conference October 2007
X-ray scatter correction method for dedicated breast computed tomography: X-ray scatter correction for dedicated breast CT journal May 2012
Image reconstruction with shift-variant filtration and its implication for noise and resolution properties in fan-beam computed tomography journal March 2003
A quality assurance program for image quality of cone-beam CT guidance in radiation therapy: Image quality QA of cone-beam CT journal April 2008
Projection x-ray imaging with photon energy weighting: experimental evaluation with a prototype detector journal July 2009
Favorable noise uniformity properties of Fourier-based interpolation and reconstruction approaches in single-slice helical computed tomography journal May 2002
Nonstationary noise characteristics of the helical scan and its impact on image quality and artifacts journal September 1997
Photon counting spectral breast CT: Effect of adaptive filtration on CT numbers, noise, and contrast to noise ratio: Effect of adaptive filtration in spectral breast CT journal April 2013
A comprehensive analysis of DgNCT coefficients for pendant-geometry cone-beam breast computed tomography journal January 2004
Spectral optimization for dedicated breast CT: Spectral optimization for dedicated breast CT journal December 2010
Effect of shaped filter design on dose and image quality in breast CT journal May 2013
A PENELOPE -based system for the automated Monte Carlo simulation of clinacs and voxelized geometries-application to far-from-axis fields : Automated MC simulation of clinacs with journal October 2011
Breast CT journal August 2007
The influence of patient centering on CT dose and image noise: How patient centering affects CT dose and noise journal June 2007
Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies journal July 2012
Experimental spectral measurements of heavy K -edge filtered beams for x-ray computed mammotomography journal January 2007
X-ray scatter characterization in dedicated breast CT with bowtie filters conference March 2014
The effects of compensator and imaging geometry on the distribution of x-ray scatter in CBCT: CBCT scatter analysis: Compensator and imaging geometry journal January 2011
The influence of bowtie filtration on cone-beam CT image quality: The influence of bowtie filtration on cone-beam CT image quality journal December 2008
Dedicated Breast CT: Radiation Dose and Image Quality Evaluation journal December 2001
Experimentally determined spectral optimization for dedicated breast computed tomography: Experimentally determined spectral optimization for dedicated breast CT journal January 2011

Similar Records

Bowtie filters for dedicated breast CT: Theory and computational implementation
Journal Article · Sun Mar 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:1401377

Bowtie filters for dedicated breast CT: Analysis of bowtie filter material selection
Journal Article · Fri Aug 14 00:00:00 EDT 2015 · Medical Physics · OSTI ID:1401377

Bowtie filters for dedicated breast CT: Analysis of bowtie filter material selection
Journal Article · Tue Sep 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:1401377

Related Subjects