Calibration of a High Resolution Soft X-ray Spectrometer
A high resolution grating spectrometer (HRGS) with 2400 line/mm variable line spacing grating for the 10-50 {angstrom} wavelength range has been designed for laser-produced plasma experiments at the Lawrence Livermore National Laboratory (LLNL). The spectrometer has a large radius of curvature, R=44.3 m, is operated at a 2{sup o} grazing angle and can record high signal-to-noise spectra when used with a low-noise, cooled, charge-coupled device detector. The instrument can be operated with a 10-25 {micro}m wide slit to achieve the best spectral resolving power on laser plasma sources, approaching 2000, or in slitless mode with a small symmetrical emission source. Results will be presented for the spectral response of the spectrometer cross-calibrated at the LLNL Electron Beam Ion Trap facility using the broadband x-ray energy EBIT Calorimeter Spectrometer (ECS).
- Research Organization:
- Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 972403
- Report Number(s):
- LLNL-PROC-423239; TRN: US1001592
- Resource Relation:
- Journal Volume: 130; Conference: Presented at: 11th International Conference on X-ray Lasers, Belfast, United Kingdom, Aug 17 - Aug 22, 2008
- Country of Publication:
- United States
- Language:
- English
Similar Records
Calibration of a high resolution grating soft x-ray spectrometer
Calibration of a high resolution grating soft x-ray spectrometer
Related Subjects
GENERAL PHYSICS
74 ATOMIC AND MOLECULAR PHYSICS
CALIBRATION
CALORIMETERS
CHARGE-COUPLED DEVICES
ELECTRON BEAMS
LASER-PRODUCED PLASMA
LASERS
LAWRENCE LIVERMORE NATIONAL LABORATORY
PLASMA
RESOLUTION
SPECTRA
SPECTRAL RESPONSE
SPECTROMETERS
WAVELENGTHS
X-RAY LASERS