Sample extraction and injection with a microscale preconcentrator.
- Advanced Sensor Technologies, Albuquerque, NM
This report details the development of a microfabricated preconcentrator that functions as a fully integrated chemical extractor-injector for a microscale gas chromatograph (GC). The device enables parts-per-billion detection and quantitative analysis of volatile organic compounds (VOCs) in indoor air with size and power advantages over macro-scale systems. The 44 mm{sup 3} preconcentrator extracts VOCs using highly adsorptive, granular forms of graphitized carbon black and carbon molecular sieves. The micron-sized silicon cavities have integrated heating and temperature sensing allowing low power, yet rapid heating to thermally desorb the collected VOCs (GC injection). The keys to device construction are a new adsorbent-solvent filling technique and solvent-tolerant wafer-level silicon-gold eutectic bonding technology. The product is the first granular adsorbent preconcentrator integrated at the wafer level. Other advantages include exhaustive VOC extraction and injection peak widths an order of magnitude narrower than predecessor prototypes. A mass transfer model, the first for any microscale preconcentrator, is developed to describe both adsorption and desorption behaviors. The physically intuitive model uses implicit and explicit finite differences to numerically solve the required partial differential equations. The model is applied to the adsorption and desorption of decane at various concentrations to extract Langmuir adsorption isotherm parameters from effluent curve measurements where properties are unknown a priori.
- Research Organization:
- Sandia National Laboratories (SNL), Albuquerque, NM, and Livermore, CA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC04-94AL85000
- OSTI ID:
- 946562
- Report Number(s):
- SAND2007-0152; TRN: US200903%%912
- Country of Publication:
- United States
- Language:
- English
Similar Records
Microfabricated planar preconcentrator
Characterizing the Performance of a Proton-Transfer-Reaction Mass Spectrometer with a Rapid Cycling Tenax Preconcentrator
Related Subjects
ORGANIC
PHYSICAL AND ANALYTICAL CHEMISTRY
47 OTHER INSTRUMENTATION
GAS CHROMATOGRAPHY
SAMPLE HOLDERS
CONCENTRATORS
MINIATURIZATION
VOLATILE MATTER
INDOOR AIR POLLUTION
ADSORPTION
DESORPTION
DECANE
ADSORPTION ISOTHERMS
Gas chromatography.
Volatile organic compounds-Analysis.
Chromatographic analysis.