skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Use of a near back-scattering imaging system on the National Ignition Facility

Conference ·
OSTI ID:899385

A near back-scattering imaging diagnostic system has been implemented, qualified and fielded on the first quad of beams on the National Ignition Facility. This diagnostic images diffusing scatter plates, placed around the final focus lenses on the NIF target chamber, to quantitatively measure the fraction of light back-scattered outside of the incident cone of the focusing optics. The imaging system consists of a wide-angle lens coupled to a gated CCD camera, providing 3mm resolution over a 2m field of view. To account for changes of the system throughput due to exposure to target debris the system was routinely calibrated in situ at 532nm and 355nm using a dedicated pulsed laser source. The diagnostic and calibration methods will be described together with recent results from the NIF early light shots.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
899385
Report Number(s):
UCRL-CONF-221190; TRN: US0701954
Resource Relation:
Conference: Presented at: High Temperature Plasma Diagnostics, Williamsburg, VA, United States, May 07 - May 11, 2006
Country of Publication:
United States
Language:
English