skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Pentaerythritol Tetranitrate (PETN) Surveillance by HPLC-MS: Instrumental Parameters Development

Technical Report ·
DOI:https://doi.org/10.2172/886931· OSTI ID:886931

Surveillance of PETN Homologs in the stockpile here at LLNL is currently carried out by high performance liquid chromatography (HPLC) with ultra violet (UV) detection. Identification of unknown chromatographic peaks with this detection scheme is severely limited. The design agency is aware of the limitations of this methodology and ordered this study to develop instrumental parameters for the use of a currently owned mass spectrometer (MS) as the detection system. The resulting procedure would be a ''drop-in'' replacement for the current surveillance method (ERD04-524). The addition of quadrupole mass spectrometry provides qualitative identification of PETN and its homologs (Petrin, DiPEHN, TriPEON, and TetraPEDN) using a LLNL generated database, while providing mass clues to the identity of unknown chromatographic peaks.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
886931
Report Number(s):
UCRL-TR-216908; TRN: US200617%%392
Country of Publication:
United States
Language:
English