Title: The Dynamic Range of Ultra-High Resolution Cryogenic Gamma-ray Spectrometers

Conference ·

We are developing high-resolution cryogenic gamma-ray spectrometers for nuclear science and non-proliferation applications. The gamma-ray detectors are composed of a bulk superconducting Sn foil absorber attached to multilayer Mo/Cu transition-edge sensors (TES). The energy resolution achieved with a 1 x 1 x 0.25 mm{sup 3} Sn absorber is 50 -90eV for {gamma}-rays up to 100 keV and it decreases for large absorber sizes. We discuss the trade-offs between energy resolution and dynamic range, as well as development of TES arrays for higher count rates and better sensitivity.

Research Organization:
Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
W-7405-ENG-48
OSTI ID:
877802
Report Number(s):
UCRL-CONF-214511; TRN: US0601696
Resource Relation:
Journal Volume: 559; Journal Issue: 2; Conference: Presented at: Low Temperature Detectors (LTD) 11, Tokyo, Japan, Jul 31 - Aug 05, 2005
Country of Publication:
United States
Language:
English

References (2)

Design of a multichannel ultra-high-resolution superconducting gamma-ray spectrometer
  • Friedrich, Stephan F.; Terracol, Stephane F.; Miyazaki, Toshiyuki
  • Optical Science and Technology, the SPIE 49th Annual Meeting, SPIE Proceedings https://doi.org/10.1117/12.563624
conference October 2004
Quantifying the benefits of ultrahigh energy resolution for gamma-ray spectrometry journal March 2005