skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Inline Ultrasonic Rheometry of a Non-Newtonian Waste Simulant

Abstract

This is a discussion of non-invasive determination of the viscosity of a non-Newtonian fluid in laminar pipe flow over the range of shear rates present in the pipe. The procedure requires knowledge of the flow profile in and the pressure drop along the long straight run of pipe. The profile is determined by using a pulsed ultrasonic Doppler velocimeter. This approach is ideal for making non-invasive, real-time measurements for monitoring and control. Rheograms of a shear thinning, thixotropic gel which is often used as a Hanford waste simulant are presented. The operating parameters and limitations of the ultrasound based instrument will be discussed. The component parts of the instrument have been packaged into a unit for field use. The presentation also discusses the features and engineering optimizations done to enhance field usability of the instrument.

Authors:
;
Publication Date:
Research Org.:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Org.:
USDOE Office of Science (SC) (US)
OSTI Identifier:
826055
Resource Type:
Conference
Resource Relation:
Conference: 227th American Chemical Society Meeting, Anaheim, CA (US), 03/28/2004--04/01/2004; Other Information: PBD: 31 Mar 2004
Country of Publication:
United States
Language:
English
Subject:
37 INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; 54 ENVIRONMENTAL SCIENCES; MONITORING; PRESSURE DROP; SHEAR; ULTRASONIC WAVES; VISCOSITY; WASTES

Citation Formats

Pfund, David M, and Pappas, Richard A. Inline Ultrasonic Rheometry of a Non-Newtonian Waste Simulant. United States: N. p., 2004. Web.
Pfund, David M, & Pappas, Richard A. Inline Ultrasonic Rheometry of a Non-Newtonian Waste Simulant. United States.
Pfund, David M, and Pappas, Richard A. 2004. "Inline Ultrasonic Rheometry of a Non-Newtonian Waste Simulant". United States. https://www.osti.gov/servlets/purl/826055.
@article{osti_826055,
title = {Inline Ultrasonic Rheometry of a Non-Newtonian Waste Simulant},
author = {Pfund, David M and Pappas, Richard A},
abstractNote = {This is a discussion of non-invasive determination of the viscosity of a non-Newtonian fluid in laminar pipe flow over the range of shear rates present in the pipe. The procedure requires knowledge of the flow profile in and the pressure drop along the long straight run of pipe. The profile is determined by using a pulsed ultrasonic Doppler velocimeter. This approach is ideal for making non-invasive, real-time measurements for monitoring and control. Rheograms of a shear thinning, thixotropic gel which is often used as a Hanford waste simulant are presented. The operating parameters and limitations of the ultrasound based instrument will be discussed. The component parts of the instrument have been packaged into a unit for field use. The presentation also discusses the features and engineering optimizations done to enhance field usability of the instrument.},
doi = {},
url = {https://www.osti.gov/biblio/826055}, journal = {},
number = ,
volume = ,
place = {United States},
year = {Wed Mar 31 00:00:00 EST 2004},
month = {Wed Mar 31 00:00:00 EST 2004}
}