Title: Monitoring system of the Tritium Research Laboratory, Sandia Laboratories, Livermore, CA

Technical Report ·
DOI:https://doi.org/10.2172/6439592· OSTI ID:6439592

Automated tritium monitoring is now in use at the Tritium Research Laboratory (TRL). Betatec 100 tritium monitors, along with several Sandia-designed accessories, have been combined with a PDP 11/40 computer to automatically read and record tritium concentrations of room air, containment, and cleanup systems. Each individual monitoring system, in addition to a local display in the area of interest, has a visible/audible display in the control room. Each system is then channeled into the PDP 11/40 computer, providing immediate assessment of the status of the entire laboratory from a central location. Measurement capability ranges from ..mu..Ci/m/sup 3/ levels for room air monitoring to kCi/m/sup 3/ levels for glove box and cleanup systems monitoring. In this report the overall monitoring system and its capabilities are discussed, with detailed descriptions given of monitors and their components.

Research Organization:
Sandia National Lab. (SNL-CA), Livermore, CA (United States)
DOE Contract Number:
EY-76-C-04-0789
OSTI ID:
6439592
Report Number(s):
SAND-78-8608; TRN: 79-004230
Country of Publication:
United States
Language:
English