Title: Simulation evaluation of NIST air-kerma rate calibration standard for electronic brachytherapy: Simulation of NIST air-kerma rate eBT calibration standard

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4940791· OSTI ID:1498033
 [1];  [2];  [3]
  1. Univ. of Massachusetts, Lowell, MA (United States)
  2. Tufts University School of Medicine, Boston, MA (United States)
  3. Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

Purpose: Dosimetry for the model S700 50 kV electronic brachytherapy (eBT) source (Xoft, Inc., a subsidiary of iCAD, San Jose, CA) was simulated using Monte Carlo (MC) methods by Rivard et al. [“Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent x-ray source: An electronic brachytherapy source,” Med. Phys. 33, 4020–4032 (2006)] and recently by Hiatt et al. [“A revised dosimetric characterization of the model S700 electronic brachytherapy source containing an anode-centering plastic insert and other components not included in the 2006 model,” Med. Phys. 42, 2764–2776 (2015)] with improved geometric characterization. Additionally, while these studies examined the dose distribution in water, there have not previously been reports of the eBT source calibration methods beyond that recently reported by Seltzer et al. [“New national air-kerma standard for low-energy electronic brachytherapy sources,” J. Res. Natl. Inst. Stand. Technol. 119, 554–574 (2014)]. Therefore, the motivation for the current study was to provide an independent determination of air-kerma rate at 50 cm in air Kair(d = 50 cm) using MC methods for the model S700 eBT source. Methods: Using CAD information provided by the vendor and disassembled sources, an MC model was created for the S700 eBT source. Simulations were run using the mcnp6 radiation transport code for the NIST Lamperti air ionization chamber according to specifications by Boutillon et al. [“Comparison of exposure standards in the 10-50 kV x-ray region,” Metrologia 5, 1–11 (1969)], in air without the Lamperti chamber, and in vacuum without the Lamperti chamber. Kair(d = 50 cm) was determined using the *F4 tally with NIST values for the mass energy-absorption coefficients for air. Photon spectra were evaluated over 2π azimuthal sampling for polar angles of 0° ≤ θ ≤ 180° every 1°. Volume averaging was averted through tight radial binning. Photon energy spectra were determined over all polar angles in both air and vacuum using the F4 tally with 0.1 keV resolution. A total of 1011 simulated histories were run for the Lamperti chamber geometry (statistical uncertainty of 0.14%), with 1010 histories for the in-air and in-vacuum simulations (statistical uncertainty of 0.04%). The total standard uncertainty in the calculated air-kerma rate determination amounted to 6.8%. Results: MC simulations determined the air-kerma rate at 50 cm from the source with the modeled Lamperti chamber to be (1.850 ± 0.126) × 10-4 Gy/s, which was within the range of Kair(d = 50 cm) values (1.67–2.11) × 10-4 Gy/s measured by NIST. The ratio of the photon spectra in air and in vacuum were in good agreement above 13 keV, and for θ < 150° where the influence of the Kovar sleeve and the Ag epoxy components caused increased scatter in air. Below 13 keV, the ratio of the photon spectra in air to vacuum exhibited a decrease that was attributed to increased attenuation of the photons in air. Across most of the energy range on the source transverse plane, there was good agreement between the authors' simulated spectra and that measured by NIST. Discrepancies were observed above 40 keV where the NIST spectrum had a steeper fall-off towards 50 keV. Conclusions: Through MC simulations of radiation transport, this study provided an independent validation of the measured air-kerma rate at 50 cm in air at NIST for the model S700 eBT source, with mean results in agreement within 3.3%. Lastly, this difference was smaller than the range (i.e., 23%) of the measured values.

Research Organization:
Los Alamos National Laboratory (LANL), Los Alamos, NM (United States)
Sponsoring Organization:
USDOE
Grant/Contract Number:
89233218CNA000001
OSTI ID:
1498033
Report Number(s):
LA-UR-15-29260
Journal Information:
Medical Physics, Vol. 43, Issue 3; ISSN 0094-2405
Publisher:
American Association of Physicists in MedicineCopyright Statement
Country of Publication:
United States
Language:
English
Citation Metrics:
Cited by: 7 works
Citation information provided by
Web of Science

References (23)

A modified dose calculation formalism for electronic brachytherapy sources journal May 2015
New National Air-Kerma Standard for Low-Energy Electronic Brachytherapy Sources
  • Seltzer, Stephen M.; O'Brien, Michelle; Mitch, Michael G.
  • Journal of Research of the National Institute of Standards and Technology, Vol. 119 https://doi.org/10.6028/jres.119.022
journal January 2014
On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry journal March 2012
Spectroscopic characterization of a novel electronic brachytherapy system journal December 2007
Initial MCNP6 Release Overview journal December 2012
A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC-ESTRO: AAPM TG-138 and GEC-ESTRO brachytherapy dosimetry uncertainty recommendations journal January 2011
Influence of photon energy spectra from brachytherapy sources on Monte Carlo simulations of kerma and dose rates in water and air: Influence of brachytherapy source photon energy spectra journal January 2010
Procedures for establishing and maintaining consistent air-kerma strength standards for low-energy, photon-emitting brachytherapy sources: Recommendations of the Calibration Laboratory Accreditation Subcommittee of the American Association of Physicists i journal February 2004
Radiobiology for eye plaque brachytherapy and evaluation of implant duration and radionuclide choice using an objective function: Radiobiological evaluation of eye plaque brachytherapy treatments journal May 2012
Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent X-Ray Source: An electronic brachytherapy sourcea): Dosimetry parameters for the Xoft Axxent X-Ray Source journal October 2006
Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations journal February 2004
A revised dosimetric characterization of the model S700 electronic brachytherapy source containing an anode-centering plastic insert and other components not included in the 2006 model: Revised dosimetric characterization of an electronic brachytherapy source journal May 2015
Anode optimization for miniature electronic brachytherapy X-ray sources using Monte Carlo and computational fluid dynamic codes journal March 2016
Evidential Value That Exercise Improves BMI z -Score in Overweight and Obese Children and Adolescents journal January 2015
Characteristics of miniature electronic brachytherapy x-ray sources based on TG-43U1 formalism using Monte Carlo simulation techniques: MC characteristics of miniature electronic brachytherapy x-ray sources journal March 2012
SU-FF-T-309: Monte Carlo Study of An Electronic Brachytherapy Source Using MCNP5 and EGSnrc journal June 2007
Comparison of Exposure Standards in the 10–50 kV X-Ray Region journal January 1969
A comparison of MCNP4C electron transport with ITS 3.0 and experiment at incident energies between 100 keV and 20 MeV: influence of voxel size, substeps and energy indexing algorithm journal April 2002
Comparison of Exposure Standards in the 10–50 kV X-Ray Region journal January 1969
Initial MCNP6 Release Overview journal December 2012
A comparison of MCNP4C electron transport with ITS 3.0 and experiment at incident energies between 100 keV and 20 MeV: influence of voxel size, substeps and energy indexing algorithm journal April 2002
On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry journal March 2012
Spectroscopic characterization of a novel electronic brachytherapy system journal December 2007

Cited By (2)

Determination of absorbed dose to water from a miniature kilovoltage x-ray source using a parallel-plate ionization chamber journal December 2017
Treatment of cervical cancer with electronic brachytherapy
  • Lozares-Cordero, Sergio; Font-Gómez, José Antonio; Gandía‐Martínez, Almudena
  • Journal of Applied Clinical Medical Physics, Vol. 20, Issue 7 https://doi.org/10.1002/acm2.12657
journal June 2019

Figures / Tables (5)