Small-angle scattering instruments on a 1 MW long pulse spallation source
Two small-angle neutron scattering instruments have been designed and optimized for installation at a 1 MW long pulse spallation source. The first of these instruments allows access to length scales in materials from 10 to 400 {angstrom}, and the second instrument from 40 to 1200 {angstrom}. Design characteristics were determined and optimization was done using the MCLIB Monte Carlo instrument simulation package. The code has been {open_quote}benchmarked{close_quote} by simulating the {open_quote}as-built{close_quote} D11 spectrometer at ILL and a performance comparison of the three instruments was made. Comparisons were made by evaluating the scattered intensity for {delta} scatterers at different Q values for various instrument configurations needed to span a Q-range of 0.0007 - 0.44 {angstrom}{sup {minus}1}.